7.
If the calibration passed, push
enter
to continue.
8.
If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to
on page 17.
9.
On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES
The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO
The sensor was calibrated previously with this controller.
10.
Reconnect the sensor, return it to the process and push
enter
. The output signal returns to the
active state and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Temperature calibration
The temperature sensor has been calibrated at the factory. However, it is recommended to always
calibrate the temperature sensor before calibrating the measurement sensor.
1.
Put the sensor in a container of water that is at a known temperature. Measure the temperature of
the water with an accurate thermometer or independent instrument.
2.
Push the
menu
key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
3.
If the pass code is enabled in the security menu for the controller, enter the pass code.
4.
Select 1 PT TEMP CAL and push
enter
.
5.
The raw temperature value is displayed. Push
enter
.
6.
Enter the correct value if different from that displayed and push
enter
.
7.
Push
enter
to confirm the calibration.
Exit calibration procedure
If the
back
key is pushed during a calibration, the user can exit the calibration.
1.
Push the
back
key during a calibration. Three options are shown:
Option
Description
QUIT CAL
Stop the calibration. A new calibration must start from the beginning.
BACK TO CAL
Return to the calibration.
LEAVE CAL
Exit the calibration temporarily. Access to other menus is allowed. To return to the
calibration, push the
menu
key and select SENSOR SETUP.
2.
Use the arrow keys to select one of the options and push
enter
.
Change calibration options
The user can set a calibration reminder or include an operator ID with calibration data from this
menu.
1.
Push the
menu
key and select SENSOR SETUP>[Select Sensor]>CALIBRATE.
2.
If the pass code is enabled in the security menu for the controller, enter the pass code.
English
17
Содержание 8332
Страница 2: ...English 3 Français 22 Español 42 中文 64 2 ...
Страница 7: ...English 7 ...
Страница 8: ...8 English ...
Страница 26: ...26 Français ...
Страница 27: ...Français 27 ...
Страница 46: ...46 Español ...
Страница 47: ...Español 47 ...
Страница 68: ...68 中文 ...
Страница 69: ...中文 69 ...
Страница 81: ......