![Hach 5410001 Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/hach/5410001/5410001_user-manual_3419512006.webp)
Specification
Details
Sample flow rate to sample
conditioning
50 to 500 mL/minute flow rate necessary (250 mL/minute
recommended)
Inlet pressure to instrument
1 to 5 psig (0.07 to 0.34 bar), 1.5 psig (0.1 bar) is optimum, >
5 psig (0.34 bar) can cause sample tubing failure
Inlet pressure to sample
conditioning
1.5 psig to 75 psig (0.1 to 5.2 bar)
Inlet fitting
At instrument, 6.35 mm (0.25 in.) OD polyethylene tubing
with quick-disconnect fitting
Drain fitting
Hose barb for 12.7 mm (0.5 in.) ID flexible tubing
Sample temperature range
5 to 40 °C (41 to 104 °F)
Certifications
CE, cETLus
Warranty
1 year (EU: 2 years)
Trip point specifications
Alarm trip point Minimum trip
value
Maximum trip value Temperature influence on trip
point
0.3 mg/L
0.22 mg/L
0.38 mg/L
–0.03 mg/L per °C
1 mg/L
0.75 mg/L
1.25 mg/L
–0.03 mg/L per °C
2 mg/L
1.5 mg/L
2.5 mg/L
–0.03 mg/L per °C
5 mg/L
3.75 mg/L
6.25 mg/L
–0.06 mg/L per °C
10 mg/L
7.5 mg/L
12.5 mg/L
–0.08 mg/L per °C
20 mg/L
15.0 mg/L
25.0 mg/L
–0.09 mg/L per °C
50 mg/L
37.5 mg/L
62.5 mg/L
–0.29 mg/L per °C
100 mg/L
75.0 mg/L
125.0 mg/L
–0.60 mg/L per °C
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product
including, without limitation, direct, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user is soley responsible to identify
critical application risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.