8.11 Replace the fuse
D A N G E R
Electrocution hazard. Remove power from the instrument before doing maintenance or
service activities.
D A N G E R
Fire hazard. Use the same type and current rating to replace fuses.
Remove the two fuses (F1 and F2) and replace them with two new fuses with the same
specifications, T, 1.25 A, 250 V. The same fuse rating is used for the 115 V and for the 230 V
operation. Refer to
on page 14.
Section 9 Troubleshooting
Problem
Possible cause
Solution
The SOFT LED light is not on
and the pump motor does not
operate.
No power
• Make sure that the power switch is
set to on.
• Connect the power cord.
• Replace the fuses if necessary.
The SOFT LED light is not on
and the pump motor does
operate.
Problem with the power
supply
Replace the main circuit board.
Contact technical support.
The SOFT LED light is on but
the pump motor does not
operate.
Operation power is low
Make sure that the line voltage is
within specification.
Voltage selector setting is
not correct
Set the line voltage selector switch to
the correct voltage.
Motor cable is not
connected
Connect the motor cable connection.
Motor does not operate
Replace the motor.
The reading is continually high.
The HARD LED is on.
• No stir bar
• LED is not connected
• No sample flow
• Out of reagents
Refer to
on page 22. If the problem
continues, cycle power to the
instrument, examine all power
connections and fuses, replace the
circuit board or contact technical
support.