8.11 Replace the fuse

D A N G E R

 

Electrocution hazard. Remove power from the instrument before doing maintenance or
service activities.

D A N G E R

 

Fire hazard. Use the same type and current rating to replace fuses.

Remove the two fuses (F1 and F2) and replace them with two new fuses with the same
specifications, T, 1.25 A, 250 V. The same fuse rating is used for the 115 V and for the 230 V
operation. Refer to 

Figure 10

 on page 14.

Section 9 Troubleshooting

Problem

Possible cause

Solution

The SOFT LED light is not on
and the pump motor does not
operate.

No power

• Make sure that the power switch is

set to on.

• Connect the power cord.
• Replace the fuses if necessary.

The SOFT LED light is not on
and the pump motor does
operate.

Problem with the power
supply

Replace the main circuit board.
Contact technical support.

The SOFT LED light is on but
the pump motor does not
operate.

Operation power is low

Make sure that the line voltage is
within specification.

Voltage selector setting is
not correct

Set the line voltage selector switch to
the correct voltage.

Motor cable is not
connected

Connect the motor cable connection.

Motor does not operate

Replace the motor.

The reading is continually high.
The HARD LED is on.

• No stir bar
• LED is not connected
• No sample flow
• Out of reagents

Refer to 

Troubleshooting for a hard

reading

 on page 22. If the problem

continues, cycle power to the
instrument, examine all power
connections and fuses, replace the
circuit board or contact technical
support.

English

   

21

Содержание 5410001

Страница 1: ...DOC023 53 80422 SP 510 03 2022 Edition 13 User Manual...

Страница 2: ......

Страница 3: ...Electrical connections for conduit 13 3 3 3 Wiring for power 14 3 3 4 Select the voltage 14 3 3 5 Connect the alarm relays 14 3 4 Install the buffer solution 15 3 5 Install the indicator solution 15...

Страница 4: ...8 11 Replace the fuse 21 Section 9 Troubleshooting 21 9 1 Troubleshooting for a hard reading 22 Section 10 Replacement parts and consumables 23 Index 25 Table of Contents 2...

Страница 5: ...ter alarm transition One reading below the trip point is necessary to trigger a hard to soft water alarm transition Trip point 0 3 1 2 5 10 20 50 and 100 mg L Refer to Trip point specifications on pag...

Страница 6: ...0 60 mg L per C Section 2 General information In no event will the manufacturer be liable for direct indirect special incidental or consequential damages resulting from any defect or omission in this...

Страница 7: ...mbol to avoid potential injury If on the instrument refer to the instruction manual for operation or safety information Electrical equipment marked with this symbol may not be disposed of in European...

Страница 8: ...e the user will be required to correct the interference at their expense The following techniques can be used to reduce interference problems 1 Disconnect the equipment from its power source to verify...

Страница 9: ...Indicator and reagent bottles 4 Air purge optional 10 Sample inlet 5 Access cover 11 Enclosure drain 6 Colorimeter 12 Sample drain 2 3 Product components Make sure that all components have been receiv...

Страница 10: ...qualified personnel must conduct the tasks described in this section of the document 3 1 Install the instrument Install this instrument on an indoor wall away from direct sunlight For a complete purge...

Страница 11: ...Figure 3 Dimensions for wall installation Figure 4 Door clearance English 9...

Страница 12: ...e off of the connector when water pressure is applied Note The sample drain connector will install on inch ID flexible tubes customer supplied 3 2 1 Connect the air purge optional An air purge is nece...

Страница 13: ...he pump tubes before installation Make sure to turn the screws in small increments and move from one screw to the other so that the plate is pulled down equally Do not overtighten To install the press...

Страница 14: ...oint 8 Low flow valve option 3 Drainage flow 6 Filtered sample bypass ball valve shown open 9 High flow valve option 3 3 Electrical installation D A N G E R Electrocution hazard Always remove power to...

Страница 15: ...ain power source W A R N I N G Electrocution hazard Install a 10 A circuit breaker for mains power Identify the circuit breaker with a label as a local disconnect for this equipment The instrument has...

Страница 16: ...strument is set to 115 volt operation at the factory To change the line voltage to 230 volt operation slide the voltage switch to the 230 V position Refer to Figure 10 on page 14 Note The fuses are ap...

Страница 17: ...nal block Terminal 1 Terminal 2 Terminal 3 J7 COM Normally open NO Normally closed NC 3 4 Install the buffer solution Buffer solutions are formulated at the factory and are ready to install without pr...

Страница 18: ...stir bar A stir bar is included in the installation kit Install the stir bar in the colorimeter sample cell for correct operation Refer to Figure 13 Figure 13 Install the stir bar Section 4 User inter...

Страница 19: ...to Install the pump valve pressure plate on page 11 1 Open the supply valve to the instrument 2 Let the pressure in the tubes stabilize Note If leaks occur under pressure examine all connections Secu...

Страница 20: ...steps 2 4 again 6 Push ZERO CAL The zero cal LED is on continuously 7 When the zero cal LED flashes add two drops of EDTA Solution 10 g L into the colorimeter 8 When the LED stops flashing and is on c...

Страница 21: ...otective equipment appropriate to the chemicals that are handled Refer to the current safety data sheets MSDS SDS for safety protocols Refer to the MSDS for instructions to safely clean reagent spills...

Страница 22: ...are handled Refer to the current safety data sheets MSDS SDS for safety protocols The colorimeter measuring cell can collect sediment or cause layers to grow on the inside walls Clean the cell with a...

Страница 23: ...es if necessary The SOFT LED light is not on and the pump motor does operate Problem with the power supply Replace the main circuit board Contact technical support The SOFT LED light is on but the pum...

Страница 24: ...block and cause the color to be diluted Make sure to set the flow rate to 200 mL minute The sample line is pinched off in the pinch block Pull the sample line from the colorimeter Water that is alrea...

Страница 25: ...ent reads hard again Section 10 Replacement parts and consumables W A R N I N G Personal injury hazard Use of non approved parts may cause personal injury damage to the instrument or equipment malfunc...

Страница 26: ...10 g L 1 oz 102133 Magnesium standard solution dropping bottle 10 g L 1 oz 102233 SP510 Hardness Monitor indicator solution for 25 ppm 500 mL 2502749 SP510 Hardness Monitor indicator solution for 50 p...

Страница 27: ...2 Colorimeter 20 D Dimensions 8 F Fuse 21 H Hard calibration 18 Hard reading 22 I Indicator solution 15 K Keypad 16 N Navigation 16 P Pressure plate 11 R Reagent 15 19 Reagent line 22 Relay 14 S Sampl...

Страница 28: ......

Страница 29: ......

Страница 30: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: