Error/Warning

Description

Solution

Please disconnect the
USB cable from your
computer.

Data storage does not
respond while connected
to the meter and the
computer.

Disconnect the USB
cable from the meter and
try sending data again.

USB module memory
full. Delete data and try
again.

Data storage is full.

1.

 Connect USB/power

module to the computer.

2.

 Download the stored

data to the computer.

3.

 Delete Data Log on

the module.

Delete Last Reading
Failed!

Error in the data storage.

Turn the meter off and
on. If the error message
still occurs, contact
technical support.

Delete Data Log failed!

Can't read data set!

Can't store data!

Can't store to the
Reading Log!

Can't store to the Verify
Cal Log!

Error storing data!

Error reading data!

Replacement parts and accessories

Replacement parts

Description

Quantity

Item no.

StablCal ampule calibration kit

1

2971205

10 NTU verification standard

100 mL

2961701

Replacement parts (continued)

Description

Quantity

Item no.

Silicone Oil

15 mL

126936

Insert, molded bottom

1

2971507

Sample cell oiling cloth

1

4707600

1" glass sample cell (10 ml ) w/cap (Turb)

pkg/6

2434706

Carrying case (includes insert)

1

2971500

Battery set, AA alkaline batteries

pkg/4

1938004 

Lamp assy

1

4653900

Blank module

1

LZV797

Rubber foot set

1

LZV821

Lamp cover (includes screws)

1

LZV822

Battery cover (includes 2 feet)

1

LZV823

Module cover

1

LZV824

Connector cover for USB/power module

1

LZV825

Connector cover for power module

1

LZV826

Lid (includes magnet)

1

LZV827

Accessories

Description

Quantity

Item no.

USB/power module (includes: universal
power supply, USB cable, instruction
sheet)

1

LZV813.99.00002

2

Power module (includes: universal power
supply, instruction sheet)

1

LZV804.99.00002

1

2

Not available in all regions

English

   

23

Содержание 2100Qis

Страница 1: ...DOC022 53 80041 2100Q and 2100Qis 12 2017 Edition 4 User Manual...

Страница 2: ......

Страница 3: ...face and navigation 7 User interface 7 Display description 8 Navigation 8 Startup 8 Turn the meter on and off 8 Change the language 8 Change the date and time 9 Standard operation 9 Use a sample ID 9...

Страница 4: ...ion standard overview 14 StablCal RapidCal calibration 15 Verification options 16 Calibration verification Verify Cal 16 Reading modes 17 Apply silicone oil to a sample cell 17 Indexing a single cell...

Страница 5: ...Data logger 500 records Specification Details Power requirement AC 100 240 V 50 60 Hz with power or USB power module 4 AA alkaline batteries Rechargeable NiMH for use with USB power module Operating c...

Страница 6: ...tached to the instrument Personal injury or damage to the instrument could occur if not observed A symbol on the instrument is referenced in the manual with a precautionary statement This is the safet...

Страница 7: ...at their expense The following techniques can be used to reduce interference problems 1 Disconnect the equipment from its power source to verify that it is or is not the source of the interference 2 I...

Страница 8: ...artment Check the battery contacts for corrosion and clean them if necessary N O T I C E When using nickel metal hydride NiMH batteries the battery icon will not indicate a full charge after freshly c...

Страница 9: ...BRATION key shows calibration screen start calibration select cal options 6 RIGHT key contextual read turbidity sample selects or confirms options opens jumps to sub menus 3 DOWN key scroll through me...

Страница 10: ...o select an option There are two ways to change options 1 Select an option from a list Use the UP and DOWN keys to select an option If check boxes are shown more than one option can be selected Push t...

Страница 11: ...s menu 2 Select create or delete a sample ID Option Description Current ID Select an ID from a list The current ID will be associated with sample data until a different ID is selected Create a New Sam...

Страница 12: ...ode Signal Average mode or in the Rapidly Settling Turbidity mode Refer to Reading modes on page 17 for more information For accurate turbidity readings use clean sample cells and remove air bubbles d...

Страница 13: ...ace Note Do not hold the instrument while making measurements 5 Gently invert and then insert the sample cell in the instrument cell compartment so the diamond or orientation mark aligns with the rais...

Страница 14: ...entire Reading Log can be deleted at once Send stored data Data can be stored and transferred to a printer computer or USB storage device The data will be formatted as an XML file Install the USB powe...

Страница 15: ...d 3 Select Security On to enable the password setting The requirement for the password entry is controlled by setting Security Options on or off Note Set the Security to Off to disable the password se...

Страница 16: ...er to Table 1 for the calibration standard overview Table 1 Calibration overview Type of calibration Required standards StablCal RapidCal 0 40 NTU 20 NTU StablCal 0 1000 NTU 20 NTU 100 NTU 800 NTU For...

Страница 17: ...urve 3 Select StablCal RapidCal from the list and push OK 4 Insert the 20 NTU StablCal Standard and close the lid Note The standard to be inserted is bordered 5 Push Read The display shows Stabilizing...

Страница 18: ...one of the following options for time interval and push OK Off 30 min Default setting 2 h 4 h 8 h 24 h Allow Defer Push Allow Defer and select Yes or No to postpone the verification due time Calibrati...

Страница 19: ...ng Turbidity reads and calculates five readings while showing intermediate results Apply silicone oil to a sample cell Sample cells and caps must be extremely clean and free from significant scratches...

Страница 20: ...ll compartment Close the lid 5 Push the UP and DOWN key to access the Reading Options and then select Indexing Sample Cell Note The instruments always stays in the last selected reading mode 6 Push Re...

Страница 21: ...Cap and store the sample cells 3 Wipe the outside of the sample cells dry with the a soft cloth Replace the battery W A R N I N G Explosion hazard An expired battery can cause hydrogen gas buildup in...

Страница 22: ...1 2 3 4 20 English...

Страница 23: ...5 6 7 8 English 21...

Страница 24: ...tion Solution Temperature too high Switch off instrument Temperature has exceeded the meter limits 60 C or 140 F Turn off the meter and let it cool down RST Average value Solids are settling too slowl...

Страница 25: ...cement parts Description Quantity Item no StablCal ampule calibration kit 1 2971205 10 NTU verification standard 100 mL 2961701 Replacement parts continued Description Quantity Item no Silicone Oil 15...

Страница 26: ...StablCal 0 5 NTU Standard 100 mL 2698042 StablCal calibration kit 100 mL 2971210 StablCal calibration kit 500 mL 2971200 Gelex secondary standards set 1 2464105 Deionized water 4 vials 27217 Filter 0...

Страница 27: ...display 8 I Indexing a sample cell 17 K keypad 7 L language 8 M menu navigation 8 meter components 5 meter specifications 3 meters troubleshooting 22 N normal reading 17 O operator ID 9 orientation m...

Страница 28: ...V verification 16 Index 26...

Страница 29: ......

Страница 30: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: