![Hach 2100Qis Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/hach/2100qis/2100qis_user-manual_3141732017.webp)
StablCal
®
RapidCal
™
calibration
1.
Push the
CALIBRATION
key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
2.
Push the
UP
and
DOWN
key to access
Cal Options and then
select Cal.Curve.
3.
Select StablCal
®
RapidCal
™
from the list
and push
OK
.
4.
Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
5.
Push
Read
. The
display shows
Stabilizing and then
shows the result.
6.
Push
Done
to
review the calibration
details.
7.
Push
Store
to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode,
refer to
Calibration
verification (Verify Cal)
on page 16.
English
15