compartment so the diamond or orientation mark aligns with the raised orientation
mark in front of the cell compartment.
1.
Push the
CALIBRATION
key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
2.
Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
3.
Push
Read
. The
display shows
Stabilizing and then
shows the result.
4.
Repeat Step 2 and 3
with the 100 NTU and
800 NTU StablCal
Standard.
Note: Push Done to
complete a 2 point
calibration.
5.
Push
Done
to
review the calibration
details.
6.
Push
Store
to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode.
Refer to
Calibration
verification (Verify Cal)
on page 16.
Turbidity measurement
W A R N I N G
Potential explosion and fire hazard. This turbidimeter is designed for water based
samples. Do not measure solvent or combustible based samples.
Readings can be taken with the Normal reading mode, Signal Average
mode or in the Rapidly Settling Turbidity mode. Refer to
Reading modes
on page 17 for more information. For accurate turbidity readings use
clean sample cells and remove air bubbles (degassing).
Measurement notes
Proper measurement techniques are important in minimizing the effects
of instrument variation, stray light and air bubbles. Use the following
measurement notes for proper measurements.
Instrument
• Make sure that the meter is placed on a level, stationary surface
during the measurement.
Note: Do not hold the meter in the hand during measurement.
• Always close the sample compartment lid during measurement,
calibration and storage.
• Remove sample cell and batteries from the instrument if the
instrument is stored for an extended time period (more than a month).
• Keep the sample compartment lid closed to prevent the entry of dust
and dirt.
Sample cells
• Always cap the sample cell to prevent spillage of the sample into the
instrument.
• Always use clean sample cells in good condition. Dirty, scratched or
damaged cells can cause inaccurate readings.
• Make sure that cold samples do not “fog” the sample cell.
• Store sample cells filled with distilled or deionized water and cap
tightly.
Measurement
10
English