86
The measurement result is shown on the display.
Execution of a time course measurement
1.
Press
TIME COURSE
.
2.
Press
OPTIONS
to adjust the settings within time course mode.
3.
Place the feed tube in the zero solution and press
ZERO
.
The zero solution is pumped in according to the selected SIP time.
Once the settle time has passed, the zero measurement is taken.
Note:
If the
PURGE START: AUTO
setting has been activated in
the Sipper options menu, the rinse agent is pumped in automatically
after measurement.
The zero measurement is shown on the display.
4.
Place the feed tube in the sample solution and press
READ
.
The sample solution is pumped in over the selected SIP time. Once
the settle time has passed, the measurement is taken.
The measurement is taken in accordance with the time course
settings. The
ZERO/READ
buttons change to
MARK/STOP
during
the time course. To abort the measurement process, press
STOP
.