English
84
The zero measurement is shown on the display.
3.
Place the feed tube in the sample solution and press
READ
.
The sample solution is pumped in over the selected SIP time. Once
the settle time has passed, the measurement is taken.
Note:
The remaining time is shown on the display. Press
CANCEL
to stop the operation.
Note:
If the
PURGE START: AUTO
setting has been activated in
the Sipper options menu, the rinse agent is pumped in automatically
after measurement.
Note:
Either deionized water or the next sample can be used as a
rinse agent.
The measurement result is shown on the display.
, an additional purge cycle of the Pour-Thru
Cell with deionized water is required between samples.
The Dual Path Length Pour-Thru Cell, 1 inch/1 cm can also be used for
the Nessler Method for nitrogen and TKN if cleaned correctly. For this,
pour a few sodium thiosulfate crystals into the cell to clean it. Rinse out
the crystals with deionized water.
The use of the Pour-Thru Cell is not possible for the tests listed in
, since chemical problems or other complications can occur. Use
the sample cell specified in the procedural instructions for these tests.
Table 1 Tests for which purging with deionized water is required
Aluminum, Aluminon
Chlorine dioxide, LR
Cobalt, PAN
Copper, Porphyrin
Hardness, Calmagite
Manganese, LR, PAN
Nickel, PAN
Nitrate, MR
Nitrate, HR
Содержание SIP 10
Страница 18: ...Български 18 ...
Страница 32: ...Čeština 32 ...
Страница 46: ...Dansk 46 ...
Страница 76: ...Ελληνικά 76 ...
Страница 90: ...English 90 ...
Страница 104: ...Español 104 ...
Страница 132: ...Hrvatski 132 ...
Страница 160: ...Italiano 160 ...
Страница 188: ...Polski 188 ...
Страница 202: ...Português 202 ...
Страница 258: ...Svenska 258 ...
Страница 272: ...Türkçe 272 ...