43
Appendix D Modbus Register Information
Table 2 Sensor Modbus Registers
Tag Name
Register # Data Type Length
R/W
Description
Main Measurement Parameter in
mg/L
40001
Float
2
R
Concentration Measurement Tag in
mg/L
pH Measurement Param.
40003
Float
2
R
pH Measurement Tag
Temperature measurement
40005
Float
2
R
Temperature measurement
Current Measurement Parameter in
µA
40007
Float
2
R
Current measurement in µA
Main Measurement Parameter in
ppm
40009
Float
2
R
Concentration Measurement Tag in
ppm
Main Measurement Parameter in ppb
40011
Float
2
R
Concentration Measurement Tag in
ppb
Main Measurement Parameter
in µg/L
40013
Float
2
R
Concentration Measurement Tag in
µg/L
Current Measurement Parameter in
nA
40015
Float
2
R
Current measurement in nA
Raw pH measurement
40017
Float
2
R
Raw pH measurement
mV Raw measurement
40019
Float
2
R
Raw ORP measurement
Raw Temperature measurement
40021
Float
2
R
Raw Temperature measurement
AutoRange Concentration in ppX
40023
Integer
1
R
Auto Ranging Tag in ppX
AutoRange Concentration in Xg/L
40024
Integer
1
R
Auto Ranging Tag in xg/L
AutoRange Current
40025
Integer
1
R
Auto Ranging redirection of nA-µA
units
Concentration Tag-based
40026
Integer
1
R
Redirection tag for concentration
ppm-mg/L units
Temperature Tag-based
40027
Integer
1
R/W
Redirection tag for temperature unit
(°C-°F)
Sensor Name[0]
40028
Integer
1
R/W
Sensor Name[0]
Sensor Name[1]
40029
Integer
1
R/W
Sensor Name[1]
Sensor Name[2]
40030
Integer
1
R/W
Sensor Name[2]
Sensor Name[3]
40031
Integer
1
R/W
Sensor Name[3]
Sensor Name[4]
40032
Integer
1
R/W
Sensor Name[4]
Sensor Name[5]
40033
Integer
1
R/W
Sensor Name[5]
Function code
40034
Integer
1
Function code
Next Step
40035
Integer
1
Next Step
Password
40036
Pass
1
R/W
Password
Serial Number[0]
40037
Integer
1
R/W
Serial Number[0]
Serial Number[1]
40038
Integer
1
R/W
Serial Number[1]
Serial Number[2]
40039
Integer
1
R/W
Serial Number[2]
Application toogle
40040
Integer
1
R/W
9184..9187 applications
Active Concentration unit
40041
Integer
1
R/W
Active concentration unit (ppm or
mg/L)
Concentration unit toogle
40042
Bit
1
R/W
Concentration unit toogle (ppm-mg/L)
Temperature unit toogle
40043
Bit
1
R/W
Temperature unit toogle (°C-°F)
Concentration offset unit
40044
Integer
1
R
Concentration offset unit (na-µA)
Compensation pH toogle
40045
Integer
1
R/W
Compensation pH toogle
(manual-auto)
pH display format toogle
40046
Bit
1
R/W
pH display format XX.X or XX.XX
Содержание 9185sc
Страница 2: ...HACH LANGE GmbH 2005 2013 All rights reserved Printed in Germany...
Страница 10: ...8 General Information...
Страница 20: ...18 Installation...
Страница 28: ...26 Operation...
Страница 32: ...30 Troubleshooting...
Страница 38: ...36 Warranty liability and complaints...
Страница 42: ...40 9185sc Theory of Operation...
Страница 44: ...42 9187sc Theory of Operation...
Страница 48: ...46 Modbus Register Information...
Страница 50: ...48 Index...