background image

15

Installation

3.5.1.2 Using the Optional pH (9184sc TFC only)

The optional pH (

Figure 2 on page 7

) is used when analysis is needed for measuring all of 

the free available chlorine (both HOCI and OCI

). See 

4.3 Sensor Setup on page 19

 fo

selecting this option using the controller during initial sensor parameter selection. 

1.

Remove the red cap from the left side of the flow cell. 

2.

Remove the O-ring from the white blank. 

3.

Gently remove the cover from the pH probe.

4.

Put the O-ring from Step 2 onto the pH probe, sliding it gently over the glass end and 
up the probe until it is flush against the red connection.

5.

Insert the newly assembled probe into the left chamber of the flow cell.

6.

Attach the keyed electrode cable.

3.6 Connecting the Sensor with the sc Controller

3.6.1 Attaching a sc Sensor with a Quick-connect Fitting

The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the 
controller (

Figure 9

). Retain the connector cap to seal the connector opening in case the 

sensor must be removed. Optional extension cables may be purchased to extend the 
sensor cable length. If the total cable length exceeds 100 m (300 ft), a termination box 
must be installed. 

Note: 

Use of a load termination box other than Cat. No. 5867000 may result in a hazard.

Figure 9

Attaching the Sensor using the Quick-connect Fitting

Содержание 9185sc

Страница 1: ...DOC023 52 00051 9184sc Chlorine 9185sc Ozone and 9187sc Chlorine Dioxide Analyzer USER MANUAL 02 2013 Edition 2A...

Страница 2: ...HACH LANGE GmbH 2005 2013 All rights reserved Printed in Germany...

Страница 3: ...Quick connect Fitting 15 3 7 Instrument and Controller Startup 16 Section 4 Operation 19 4 1 Using the sc Controller 19 4 2 Sensor Data Logging 19 4 3 Sensor Setup 19 4 3 1 Changing the Sensor Name a...

Страница 4: ...4 Extension Cables 32 Section 8 Contact 33 Section 9 Warranty liability and complaints 35 Appendix A 9184sc Theory of Operation 37 A 1 Theory of Operation 37 A 1 1 Principle of Operation 37 Appendix...

Страница 5: ...pm 0 20 mg L HOCl 0 2 ppm 0 2 mg L O3 0 2 ppm 0 2 mg L ClO2 Detection Limit 5 ppb 0 005 mg L HOCl 5 ppb 0 005 mg L O3 10 ppb 0 01 mg L ClO2 Accuracy 2 or 10 ppb HOCl whichever is greater 3 or 10 ppb O...

Страница 6: ...0 to 60 C 4 to 140 F Operating Temperature Range 0 to 45 C 32 to 113 F Relative Humidity 10 to 90 non condensing Operating Humidity 0 to 90 non condensing Compliance The sc analyzer and sensor combina...

Страница 7: ...l systems after 12 August of 2005 In conformity with European local and national regulations European electrical equipment users must now return old or end of life equipment to the Producer for dispos...

Страница 8: ...Theory of Operation Refer to the following Appendices Appendix A 9184sc Theory of Operation on page 37 Appendix B 9185sc Theory of Operation on page 39 Appendix C 9187sc Theory of Operation on page 41...

Страница 9: ...General Instrument Schematic 1 pH Probe 9184sc only 5 Gateway behind Mounting Plate 2 pH Cell Cap 9184sc only 6 Connector 3 Connector 7 Cell Cap 4 Cable to Controller 8 Probe Body See Replacement Par...

Страница 10: ...8 General Information...

Страница 11: ...ce traveled by the sample the faster the instrument can respond and indicate changes in sample concentration The 152 mm 6 inch clearance at the bottom of the instrument is not necessary if using the a...

Страница 12: ...point is important for optimum instrument performance Figure 4 The sample analyzed must be representative of the condition of the entire system Install sample line taps into the side of larger proces...

Страница 13: ...nded when contact with chemicals is possible 1 Screw the membrane cap onto the probe body Figure 6 and Figure 7 Be careful not to touch or damage the membrane surface 2 Remove the filling screw from t...

Страница 14: ...hands and rinse sensor to remove excess electrolyte 10 Place the probe into the Flow Thru Assembly Figure 6 Probe Assembly 1 Membrane Cap to Probe Body 4 Seat Electrode into Probe Body 2 Syringe Fille...

Страница 15: ...assembled probe into the right chamber of the flow cell 3 Gently screw the retaining nut back on ensuring it is snug but do not overtighten 4 Attach the keyed electrode 1 Displays the correct way to...

Страница 16: ...nstallation Figure 8 Placing the Probe Into the Flow Thru Assembly 1 Electrode Cable Connector 3 Probe Assembly 2 Probe Retaining Nut 4 Flow Thru Assembly See Replacement Parts and Accessories on page...

Страница 17: ...il it is flush against the red connection 5 Insert the newly assembled probe into the left chamber of the flow cell 6 Attach the keyed electrode cable 3 6 Connecting the Sensor with the sc Controller...

Страница 18: ...or leaks at this time If leaks occur fix by ensuring all connections are snug but not overtightened 4 Adjust the flow regulator on the flow cell until the water begins to drain out the left drain port...

Страница 19: ...17 Installation Figure 11 Setting the Flow Rate 1 Sample inlet tubing 4 Sample 2 Flow meter adjustment knob 5 Drain tubing 3 Sample level overflow indicates the correct water level...

Страница 20: ...18 Installation...

Страница 21: ...en purchased Parameter choices depending on the instrument are Chlorine HOCL does not include pH measurement Chlorine Acid which is HOCL plus the acid verification accessory does not include pH measur...

Страница 22: ...ELECTRICAL or CHEMICAL For CHEMICAL use a sample that does not contain any oxidants Ensure that the sample source has a sufficient flow rate and that the sample is adequately mixed CAL DELAY can adju...

Страница 23: ...m pH allowed value An higher value will display a PH TOO HIGH error message LOG SETUP Allows user to select data logging interval for the sensor and temperature FILTER Adjust for s This averages the s...

Страница 24: ...pH 4 6 2 1 Process pH 1 Point Sample 1 From the Main Menu select SENSOR SETUP and confirm 2 Highlight the appropriate sensor if more than one is attached and confirm 3 Select CALIBRATE and confirm 4 S...

Страница 25: ...djust the SAMPLE VALUE X XX pH to the known pH with the keypad and confirm 9 MOVE CLEAN PROBE TO SAMPLE2 Press ENTER to Continue is displayed Confirm 10 VALUE XX XX pH TEMP XXX C is displayed Confirm...

Страница 26: ...te Refer to section 4 6 3 on page 23 If adjusting the TFC value use the Total Free Chlorine Method 9 COMPLETE SLOPE nA MG OFFSET uA is displayed and confirm to continue 10 RETURN PROBE TO PROCESS is d...

Страница 27: ...ensor is attached and confirm 3 Select CALIBRATE and confirm 4 Select CAL CONFIG and confirm 5 OUTPUT MODE displays Use the keypad to choose one of the following ACTIVE HOLD TRANSFER or CHOICE and con...

Страница 28: ...26 Operation...

Страница 29: ...ng conditions or more frequently as experience dictates Figure 12 1 Shut off the sample supply Remove the probe cable 2 Unscrew the probe retaining ring Remove the sensor Note Removing the sensor may...

Страница 30: ...e membrane Refer to section 3 5 1 on page 11 to replace the electrolyte 5 2 4 Replacing the pH Electrode 9184sc only Replace the electrode 12 to 18 months depending on the application 1 Probe Chamber...

Страница 31: ...fy the RAW value TEMP TOO HIGH Verify the actual temperature is not higher than 45 C Connect the sensor simulator and verify the RAW value RAW MEASURMENT Change the preamplifier PH TOO LOW 9184sc only...

Страница 32: ...30 Troubleshooting...

Страница 33: ...001 Replacement electrode for the 9185 sc 09185 A 1000 Replacement electrode for the 9187 sc 09184 A 1001 Replacement probe body 09184 C 4100 Filling screw 09184 C 1030 Syringe 560150 21957 Pre assemb...

Страница 34: ...on Cables Description Catalog Number Cable sensor extension 0 35 m LZX847 Cable sensor extension 5 m LZX848 Cable sensor extension 10 m LZX849 Cable sensor extension 15 m LZX850 Cable sensor extension...

Страница 35: ...0 15 41 61 20 info hach lange be www hach lange be DR LANGE NEDERLAND B V Laan van Westroijen 2a NL 4003 AZ Tiel Tel 31 0 344 63 11 30 Fax 31 0 344 63 11 50 info hach lange nl www hach lange nl HACH L...

Страница 36: ...000 Vara din Tel 385 0 42 305 086 Fax 385 0 42 305 087 info hach lange hr www hach lange hr HACH LANGE MAROC SARLAU Villa 14 Rue 2 Casa Plaisance Quartier Racine Extension MA Casablanca 20000 T l 212...

Страница 37: ...n of the fault If the customer fails to notify the supplier the product is considered approved despite the defect Further liability for any direct or indirect damages is not accepted If instrument spe...

Страница 38: ...36 Warranty liability and complaints...

Страница 39: ...xist to mention the different species of chlorine Active Chlorine HOCl hypochlorous acid It is the most powerful disinfectant up to 100 times more efficient than hypochlorite Total Free Chlorine TFC H...

Страница 40: ...ere HOCl is reduced according to the reaction At the silver electrode anode the silver is oxidized to Ag The reduction in HOCl at the cathode generates a current that is directly proportional to its p...

Страница 41: ...solubility of ozone in water decreases rapidly with temperature Effects of pH Ozone reacting with OH hydroxide ions the greater the number of these ions high pH the greater the degree of breakdown Con...

Страница 42: ...40 9185sc Theory of Operation...

Страница 43: ...iffusion of the chlorine dioxide molecules through a membrane The molecules of chlorine dioxide contained in the sample diffuse through the membrane and are then found in an electrolyte zone of very s...

Страница 44: ...42 9187sc Theory of Operation...

Страница 45: ...Integer 1 R Auto Ranging redirection of nA A units Concentration Tag based 40026 Integer 1 R Redirection tag for concentration ppm mg L units Temperature Tag based 40027 Integer 1 R W Redirection tag...

Страница 46: ...easurement 40078 Float 2 R Internal use Cal TFC Measurement 40080 Float 2 R Internal use Output Mode 40082 Integer 1 R Internal use Software version 40083 Float 2 R Software version Serial Number Stri...

Страница 47: ...n 40106 Float 2 R Internal use Measurement Logging Interval 40108 Integer 1 R W Sensor Data logging interval Temperature Logging Interval 40109 Integer 1 R W Temperature logging interval Table 2 Senso...

Страница 48: ...46 Modbus Register Information...

Страница 49: ...t Specifications 3 M Mounting Installation Considerations 10 O Optional pH 15 P Probe Components 11 R Replacement Parts and Accessories 31 S Safety 5 Sample Line 10 sc100 Warning Messages 29 Sensor Da...

Страница 50: ...48 Index...

Отзывы: