![Haag-Streit OCTOPUS 900 Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/haag-streit/octopus-900/octopus-900_instructions-for-use-manual_575796011.webp)
11
DEUTSCH
ENGLISH
FRANÇAIS
ITALIANO
ESPAÑOL
NEDERLANDS
PORTUGUÊS
SVENSKA
DEUTSCH
ENGLISH
FRANÇAIS
ITALIANO
ESPAÑOL
NEDERLANDS
PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS Doc. no. 1500.7220343.04070 – 7. Edition / 2016 – 11
** If the control unit (notebook, PC) and printer are more than 1.5 m away from the
Octopus 900, then the safety isolating transformer can be dispensed with in accor-
dance with EN 60601-1.
Neither a safety isolating transformer nor a distance of > 1.5 m from the Octopus
900 is required if a medically approved control unit or a control unit with medical
-
ly approved power supply unit without printer and without optional LAN connection
is being operated. For safety reasons, it is recommended to maintain a distance
of > 1.5 m if at all possible. Otherwise, all non-medical devices must be operated
through a safety isolating transformer.
4.2 System variant II, Octopus 900 with
PC and monitor as control unit
** If the control unit (notebook, PC) and printer are more than 1.5 m away from the
Octopus 900, then the safety isolating transformer can be dispensed with in accor-
dance with EN 60601-1.
*
**
Mains connection
Mains connection
Ethernet cable
LAN (optional)
Ethernet cable *
< 1.5 m **
PC / control unit
Control unit
USB cable
safety isolating transformer
LAN connection
Network
switch
01-IFU_Octopus900-7220343_04070_eng.indd 11
18.11.2016 14:03:44