Function of the third electrode
The instrument is equipped with a sensing electrode as standard, which is located in the upper part of the measuring tube
section, and is used to indicate flooding of the measuring tube section by liquid.
This third electrode can work in three modes.
a) If the measuring tube section is empty or not entirely flooded, i.e., the electrode is not in contact with liquid,
measurement is locked and instrument displays zero flow rate. At the same time, character Z is displayed in the lower-left
corner of the display indicating that the lock function has been enabled.
1. Select the
MEASUREMENT
option from the main menu and confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select the
LOCK
command under the option
EMPTY PIPE
in order to activate the above described function.
NOTE:
The above described function is set by default in the factory and does not need to be set in most cases.
b) If the measuring tube section is empty or not entirely flooded, i.e., the electrode is not in contact with the liquid,
measurement is not locked and this situation is only indicated. The instrument displays flow rate and, at the same time,
character Z is displayed in the lower-left corner of the display to indicate that this function has been enabled.
1. Select the
MEASUREMENT
option from the main menu and confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select the
ALARM
command under the option
EMPTY PIPE
to activate the above described function.
c) Function of the third electrode can be completely disabled.
1. Select the
MEASUREMENT
option from the main menu and confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select command
OFF
under the option
EMPTY PIPE
to activate the above described function.
I.
Password setting
A password can be set to prevent unauthorized persons from gaining access to the configuration menu.
1. Use the cursor keys to select the
PASSWORD
option from the main menu. Confirm your selection by pressing the
EXE
button. A sub-menu will be displayed.
2. Select the
NEW
command and press the
EXE
key to enable password editing. The password may consist of up to eight
characters. Use the cursor keys to set your password and confirm by pressing the
EXE
button.
3. Select the
RETURN
option to complete the set up.
The instrument will ask if you want to save the new parameters by displaying the text
SAVE SETUP?
. Select
YES
to confirm
and the flow meter will save the new parameters and switch to normal display mode. If
NO
is selected, parameters will not be
saved.
The instrument will switch to display mode and work with the new parameters until it is switched off. When the instrument is
switched on again, it will use the original/default parameters.