background image

operation 

file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc 

© 2013 h/p/cosmos sports & medical gmbh     author: fh     [email protected]     created 07.02.2013     printed 04.03.2013     page: 25 of 197 

 

 

 

N

 

Holding handrails may affect the exercise results (e.g. heart rate, ECG, oxygen uptake, etc.). 

N

 

Do not deposit any clothes, towels, jewellery, etc. on the running machine. Objects on the belt 
may be tripped over or caught in the gap during rotation of the belt. 

 
Healthy subjects should hold the handrails only in case of emergency or for weight relief. The patient is not allowed to 
jump or turn around on the rotating belt. Furthermore, he/she should be informed about emergency procedures. 

 
Depending on the prescribed or intended test or training, further devices may be 
necessary (e.g. heart rate transmitter, ECG or ergospirometry device). The doctor 
and/or the medical staff are to supervise the patient during the complete test or 
training. If the heart rate transmitter is used, adjust the belt length so that the belt 
fits tightly but is not confining. The belt should not loosen while exercising. Close 
the belt with the transmitter placed outwards (POLAR logo in right position). In 

order to allow optimal skin contact, the skin should be damp. Contact gel, as used for ECG, is an excellent moisturizer. 
Moisten the two electrodes and the skin with water or ECG contact gel. Place the transmitter so that it is directly below 
the pectoral muscle (chest), as shown in the illustration. For approximately 95 - 98 % of the subjects the "normal" 
placement of the transmitter belt is recommended. 
 

 

For further information, see chapter 5.9.4 entitled "Cardio mode 

 
The user selects one of the operation modes (manual, profile, cardio or test mode) of the remote control firmware and 
selects the prescribed test/training. The software asks the user to confirm or to change default settings. Current values 
(e.g. exercise time, distance and index) are displayed. A break can be selected with the “-“ key. “PAUS” is then 
displayed in the speed display. All current values are sustained and will be continued after restart. The test can be 
cancelled with the “STOP” key. At the end of the treadmill training data can be stored and/or printed, depending on 
selected mode, connected devices and the software version. 

 

Содержание 150/50

Страница 1: ...tos med h p cosmos locomotion 150 50 E DE med h p cosmos gaitway II F S incl lt models h p cosmos quasar quasar med h p cosmos stellar stellar med h p cosmos pulsar h p cosmos locomotion 190 65 E DE m...

Страница 2: ...or accessories can result in invalidity of this manual In case of alterations of the device or the additional equipment the latest version of the manual or the corresponding additional information al...

Страница 3: ...technicians The customer user should perform this maintenance and monitoring work Some safety checks or monitoring for examples harnesses and ropes running belt condition and position etc have to be p...

Страница 4: ...tch on procedure 26 5 2 1 Switch on procedure model range 150 50 up to 190 65 26 5 2 2 Switch on procedure model range 200 75 up to 450 300 venus saturn 27 5 3 Switch off procedure 28 5 3 1 Switch off...

Страница 5: ...device software classification 83 7 4 Installation commissioning 84 7 4 1 Unpacking and packaging 84 7 4 2 Transport 85 7 4 3 Storage 87 7 4 4 Environmental requirements room configuration 87 7 4 5 Me...

Страница 6: ...urrent 112 8 4 Lubrication of the running belt running surface 112 8 4 1 Devices with UserTerminal and without automatic oil pump 113 8 4 2 Lubrication film test 114 8 4 3 Devices without UserTerminal...

Страница 7: ...interference 126 9 6 Voltage on the device housing electric shock 127 9 7 POLAR heart rate measurement system 127 9 8 RS232 interface 128 9 8 1 Troubleshooting and testing of the RS232 interface 128 9...

Страница 8: ...Retrofitting UserTerminals laptop for external control unit 171 12 18 Retrofitting emergency stop buttons 172 12 19 Safety multiple socket 172 12 20 RS232 interface USB converter 173 12 21 POLAR hear...

Страница 9: ...osmos mercury med h p cosmos Kistler Gaitway II F h p cosmos Kistler Gaitway II S h p cosmos stratos lt h p cosmos stratos h p cosmos mercury lt h p cosmos mercury h p cosmos stellar lt med h p cosmos...

Страница 10: ...hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 10 of 197 h p cosmo...

Страница 11: ...weight support system and the gait can be analysed and corrected under safe and reproducible conditions Duration of use of the medical versions is approx 5 to 25 minutes per patient per day depending...

Страница 12: ...rol of the treadmill loading parameters and for maintenance issues Software products such as h p cosmos para graphics h p cosmos para analysis and h p cosmos para motion can be optionally applied for...

Страница 13: ...ess An additional emergency stop button can be retro fitted at all h p cosmos running machines It can for example be duplicated on the opposite handrail h p cosmos conforms with the norm EN 957 6 The...

Страница 14: ...y the stress device but does not perform diagnosis N Therapy of orthopaedic problems e g arthritis in foot ankle femoropatellar syndrome knee joint problems anterior cruciate ligament plastic N Gait a...

Страница 15: ...ic dissection N Acute coronary syndrome N Acute phlebothrombosis of the lower extremities N Febrile infections N Pregnancy N Acute thrombosis N Fresh wounds e g after surgery N Acute fracture N Damage...

Страница 16: ...restrictions be excluded by contraindications The risk of falling and consequential injuries is reduced by the safety equipment and application of a fall prevention system Remaining risks in ergometry...

Страница 17: ...el must stay in permanent patient environment of 1 5 m N Recommendation and application of weight support system and or fall stop device such as the h p cosmos safety arch with harness and automatic s...

Страница 18: ...ter entitled Contraindications N The automatic operation of the running machine modes profile cardio test remote control via PC and peripheral devices is forbidden if the health and condition of the t...

Страница 19: ...he running machine and the accessories without supervisory staff and introduction N Holding the handrails during exercise may affect the exercise results e g heart rate ECG oxygen uptake etc Under nor...

Страница 20: ...hich have an extra thick running belt A safety arch with chest belt system is obligatory in this case N In case of reverse belt rotation always use the pull cord safety stop actuator or better the saf...

Страница 21: ...erse belt rotation or use the pull cord safety stop actuator N during all use with wheels cycling wheelchair inline skating or roller ski for the r models N in case of severe danger N subjects with lo...

Страница 22: ...tion to the ECG manufacturer s manual The summary of clinical data and the risk management of h p cosmos does NOT cover any measured parameters and or vital functions which are measured and or detecte...

Страница 23: ...mos strongly recommends keeping the standard setting of this option Stop means stop for all movements and should not activate the running machine in any way It is only allowed to change option 41 afte...

Страница 24: ...with chest belt system and stop function unweighting systems should be adjusted so that the subject cannot use the rear 30 of the running deck length and the automatic stop function stops the running...

Страница 25: ...its tightly but is not confining The belt should not loosen while exercising Close the belt with the transmitter placed outwards POLAR logo in right position In order to allow optimal skin contact the...

Страница 26: ...that the expulsion fuse at the front of the device front terminal below the hood is switched on N If the running machine is at an angle of elevation 0 when switched on it will automatically move to th...

Страница 27: ...on the display Expulsion fuse and main switch 2 or 3 pole ON OFF button with control lamp ON OFF button until 2003 5 2 2 Switch on procedure model range 200 75 up to 450 300 venus saturn no Illustrati...

Страница 28: ...ture The main switch can be secured against unauthorized access with a padlock 05 After a delay time green light is blinking of 1 minute the green light stops blinking and stays illuminated Now the ru...

Страница 29: ...main switch can be secured against unauthorized access with a padlock N The ON OFF intervals must not be shorter than 1 2 minutes for models with 3 phase connection 2 3 minutes This may otherwise lead...

Страница 30: ...clination system N display keyboard and interface inactive interrupted during cut off from main power N wait 1 minute N turn palm switch to the right N push ON button at main power switch N press STAR...

Страница 31: ...nd inclination system N display active shows PULL STOP N interface active but no start speed or elevation signal executed N keyboard inactive NOT interrupted during PULL STOP stage N reconnect magnet...

Страница 32: ...rminal TouchPanel retro fit kit If your running machine has no UserTerminal no display no keyboard it can only be controlled via interface RS 232 e g via ECG Ergospirometry PC with software h p cosmos...

Страница 33: ...ay consists of six four digit LCD displays which show speed time index MET energy consumption wattage alternating distance elevation inclination and heart rate Error codes and service information can...

Страница 34: ...LED status unit definition max m min km h m s mph m min km h m s mph 0 1 0 1 or 0 01 0 1 or 0 01 0 1 or 0 01 running distance LED status unit definition m km miles m km miles 0 1 0 1 0 1 running time...

Страница 35: ...witch is blue lighted belt alignment adjust it Running belt is moving outside tolerance range yellow lighted belt out of tolerance yellow blinking belt severe out of tolerance check power supply Power...

Страница 36: ...f the screen The TouchPanel PC turns on automatically with the external control unit In case the TouchPanel PC does not boot up it can be switched on manually Therefore the external control unit has t...

Страница 37: ...is also able to control the machine with h p cosmos para control Furthermore there are a lot of additional functions like next target speed with a click on the speed display next target elevation wit...

Страница 38: ...logie Herz und Kreislaufforschung e V http leitlinien dgk org American Heart Association http www heart org and http circ ahajournals org content 96 1 345 full 5 7 Application Gait analysis and rehabi...

Страница 39: ...h p cosmos gaitway with KISTLER force plates After four weeks of gait therapy you can see a considerable improvement of gait symmetry documented by comparison of the ground reaction forces Jumper s k...

Страница 40: ...ed to the serial interface RS 232 or and using the software h p cosmos para graphics on an external PC N The treadmill belt will start moving automatically and the speed and elevation will be changed...

Страница 41: ...n any time from 3 to 131 seconds Seven different acceleration and deceleration levels are available for all modes and for remote control via interface and h p cosmos coscom protocol see the following...

Страница 42: ...acceleration level 1 or 2 is for rehabilitation and for fitness for beginners N Medium acceleration level 3 or 4 is for well trained subjects for speeding up or slowing down relatively fast N High ac...

Страница 43: ...ation in the optional functions indicates 65 flashing for 65 kg body weight weight is illuminated Enter the actual body weight with a or d then c for an accurate calculation of the POWER and ENERGY co...

Страница 44: ...speed maximum elevation and the duration of the six profiles can be scaled in order to provide 6 x 6 x 6 216 variations of profiles Standard set up is without scaling for quicker and easier access Pr...

Страница 45: ...essive interval m sec km h min Warming up 2 8 10 1 4 0 0 0 High 1 3 2 11 5 2 0 0 0 Low 1 2 8 10 1 2 0 0 0 High 2 3 6 13 0 1 0 0 0 Low 2 2 8 10 1 2 0 0 0 High 3 4 0 14 4 1 0 0 0 Low 3 2 8 10 1 2 0 0 0...

Страница 46: ...1 2 0 7 2 2 0 10 0 High 2 2 5 9 0 3 0 0 0 Low 2 2 0 7 2 2 0 10 0 High 3 2 5 9 0 3 0 0 0 Low 3 2 0 7 2 2 0 10 0 High 4 2 5 9 0 3 0 0 0 Low 4 2 0 7 2 2 0 10 0 25 min Profile 6 Progressive interval elev...

Страница 47: ...d profile no 1 6 04 Start the selected profile no 1 6 c A standard in case of deactivation in the optional functions profile is illuminated and step are illuminated alternately kJ and Watt are illumin...

Страница 48: ...press several times and then hold Example to select acceleration or deceleration level 3 press the corresponding key 3 times and then hold Standard value 4 maximum 7 levels may be limited by optional...

Страница 49: ...tely N Stop training immediately if you feel sick or dizzy pain or in case of any other health problem and consult your doctor N In case of detected or suspected interference of the wireless heart rat...

Страница 50: ...unning machine adjusts speed and elevation according to the following matrix Chart for heart rate controlled work load at the heart rate lower level Heart rate difference actual set value Speed km h E...

Страница 51: ...edical doctor you can must change this level If no changes are required confirm the value with c 07 Change upper level of heart rate aor d is illuminated flashing number indicates max level of the hea...

Страница 52: ...ubject to control the load by means of speed without elevation If speed is being reduced to 0 PAUS is displayed Current values will be sustained 02 Pause mode Interfere with the cardio mode a indicate...

Страница 53: ...COOPER test etc are endurance tests max load and max heart rate tests and should only be performed after consulting a medical doctor and under supervision of trained staff N When using the treadmill...

Страница 54: ...ited by the maximum acceleration level selected in the optional settings standard value 4 During deceleration this function alternates the degree of deceleration 06 Switch to the next or previous mode...

Страница 55: ...elect age 30 years aor d Age is illuminated indicates age 30 flashing 07 Confirm 30 years c Weight is illuminated indicates weight 65 flashing 08 Select weight 55 kg aor d Weight is illuminated indica...

Страница 56: ...4 adjustable from 1 to 5 N Break time 30 sec can be changed STOP must be activated manually by the medical doctor In an already pre defined step test it is possible to set the pre programmed pause ti...

Страница 57: ...the condition of the subject N Circuit lap length 200 m can be changed N Increment 0 5 km h can be changed STOP must be activated manually by the medical doctor when the subject is completely exhaust...

Страница 58: ...speed is increased by 0 32 km h every minute STOP must be activated manually by the medical doctor when the subject is completely exhausted 8 Ellestad A protocol e g for ECG stress test Step Duration...

Страница 59: ...to zero After completing the test the results can be printed on a host printer if connected Pre test phase patient stands on the footboards not on the belt Step Duration min sec Speed km h Elevation T...

Страница 60: ...500 m 19 0 10 1 15 min 0 0 72 Training Interval 400 m 16 0 21 0 km h Step Distance time Speed km h Elevation Acceleration level 1 400 m 16 0 0 3 2 1 00 min 0 0 3 400 m 17 0 4 1 00 min 0 0 5 400 m 18 0...

Страница 61: ...3 2 1 30 min 0 0 3 300 m 21 0 4 0 45 min 0 0 5 600 m 15 0 6 1 30 min 0 0 7 300 m 21 0 8 0 45 min 0 0 9 600 m 15 0 10 1 30 min 0 0 11 300 m 21 0 12 0 45 min 0 0 13 600 m 15 0 14 1 30 min 0 0 15 300 m...

Страница 62: ...00 5 5 8 1 00 6 1 9 1 00 6 7 10 1 00 7 3 11 1 00 7 9 12 1 00 8 5 13 3 00 1 8 Tests no 77 79 are reserved for the following extensions and up dates 80 Profile VO2 10 k Step Time min Speed km h Elevati...

Страница 63: ...00 12 0 10 8 1 00 13 0 10 9 2 00 13 0 11 10 2 00 13 0 12 83 Not engaged reserved for following extensions and up dates 84 Profile VO2 14 k Step Time min Speed km h Elevation Acceleration level 1 2 00...

Страница 64: ...eleration level 1 3 00 7 5 0 1 2 7 5 3 3 7 5 6 4 8 0 6 5 8 0 9 6 8 0 12 7 8 0 15 8 8 0 18 9 8 0 21 10 8 0 24 91 Test profile Step Time min Speed km h Elevation Acceleration level 1 3 00 9 0 0 1 2 0 30...

Страница 65: ...Test profile Step Time min Speed km h Elevation Acceleration level 1 3 00 9 0 2 1 2 0 30 0 0 2 3 3 00 10 8 2 4 0 30 0 0 2 5 3 00 12 6 2 6 0 30 0 0 2 7 3 00 14 4 2 8 0 30 0 0 4 9 3 00 14 4 4 10 0 30 0...

Страница 66: ...9 1 00 14 4 0 1 10 1 00 10 1 0 3 11 2 30 10 1 3 3 12 2 20 7 0 9 1 13 0 10 9 2 5 1 14 1 00 15 1 0 1 15 1 00 10 8 0 3 16 2 30 10 8 3 3 17 2 20 7 7 9 1 18 0 10 9 9 5 1 19 1 00 15 8 0 1 20 1 00 11 5 0 3...

Страница 67: ...control all functions of the running machine via host PC and also to serve for on line monitoring of all data During programming the displays indicate the values of the present program step and not th...

Страница 68: ...shing indicates program step St 1 05 Select speed 5 0 km h aor d 06 Confirm speed c indicates distance 0 is flashing 07 Confirm distance 0 this step is programmed after time c indicates time minutes 0...

Страница 69: ...t time 4 minutes aor d 26 Confirm time c indicates time seconds 4 00 is flashing 27 Confirm time 00 seconds c indicates elevation 0 0 is flashing 28 Select elevation 5 aor d 29 Confirm elevation 5 c i...

Страница 70: ...e 5 minutes aor d 42 Confirm time c indicates time seconds 5 00 is flashing 43 Confirm time 00 seconds c indicates elevation 0 0 is flashing 44 Confirm elevation 0 c indicates acceleration level Acc 1...

Страница 71: ...ponse Display 01 Select option mode OPxx for users Note The option mode for administrators is only accessible for authorised service technicians not for users a and simultaneously d and simultaneously...

Страница 72: ...ngineer The following intervals are set back by this operation E 01 Oil interval A OP 35 E 02 Service time interval A OP 37 E 02 Service distance interval A OP 38 OP02 Total distance covered km and in...

Страница 73: ...isplay of distance indicates unit for distance 0 km 1 miles 2 m m km or miles is blinking OP14 Unit for angle of elevation indicates Unit for elevation 0 per cent 1 degree or blinking OP15 Subject s b...

Страница 74: ...be used for new installations and configurations of medical devices and medical applications from 21 03 2010 h p cosmos para control and h p cosmos para graphics exclusively work with the h p cosmos c...

Страница 75: ...lected if the safety of the subject is provided for by the use of the fall prevention e g by the safety arch The maximum acceleration and deceleration level is NOT valid for control and operation via...

Страница 76: ...n or automatically after 2 minutes after having pressed STOP Default ON Display values will be continued added after pressing START again and will not be automatically deleted by pressing STOP Display...

Страница 77: ...10 250 kg OP 16 Bodyweight request before manual automatic start OFF no request OFF ON OP 17 Unit of energy consumption JOUL kJoule JOUL kJoule CALO kcal OP 18 Maximum speed default in cardio mode 6 0...

Страница 78: ...sar lt h p cosmos quasar Application The running machines designed for use in sports and fitness are not qualified for medical application N It is not allowed to utilize sports grade treadmills for me...

Страница 79: ...is included in the delivery In order to be able to supply you with the latest technical information and service it is important for you to fill out the form Therefore please fill out the form for reg...

Страница 80: ...e plug socket is prohibited N An overload or voltage drop even temporary of more than 20 of the mains voltage might cause malfunctions and or defects and might totally switch off the running machine I...

Страница 81: ...care not to damage the power cord A damaged cord may cause electric shock or fire N Disconnect the power plug when the unit is not used for long periods Keeping the connection may cause electric shock...

Страница 82: ...hor fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 82 of 197 7 2 Electrical safety features Potential isolation transformer Potential equalization connector Shielded motor cable Fr...

Страница 83: ...components For this reason the worst case malfunction scenario has to be taken into account h p cosmos classifies the internal firmware and software components as class B due to a number of risk redu...

Страница 84: ...the machine and accessories read the instructions on the crate Make sure that the machine power connection cable or any optional equipment will not be damaged during unpacking Pay special attention t...

Страница 85: ...ixed with a special transport fuse construction available at h p cosmos in order to avoid damage during transport Provided the right handrail is flat and temporarily fixed on the running surface the w...

Страница 86: ...p cosmos com created 07 02 2013 printed 04 03 2013 page 86 of 197 Transport with trolleys For easy transport put two normal trolleys under the device platform approximately 30 cm x 30 cm with 4 manoe...

Страница 87: ...mm x B 1490 mm 5 066 m platform Net static weight of running machine 1300 kg static body weight of subject 200 kg dynamical weight of subject 1200 kg up to 6 times the body weight total weight load of...

Страница 88: ...ject 1 must never be touched by the doctor or trainer 2 at the same time Casing and cover Parts of non medical electric devices within the surrounding of the subject which after having taken off the c...

Страница 89: ...The levelling sockets and the front wheels of the elevation system may cause a print in the flooring Please use h p cosmos floor mats to protect sensitive flooring Floor protection mat 150 50 for runn...

Страница 90: ...onsibility to check the correct function of the mains connection including the outlets An authorized electrical engineer must check these points for perfect functioning regularly 1 4 years Inspections...

Страница 91: ...our electrical engineer or h p cosmos if you have further questions N Before installing the running machine please compare the specifications on the name plate concerning the mains voltage and the mai...

Страница 92: ...sponding plug pin next to the main switch on the front and then connect the mains plug N During electric safety measurements and test leakage currents etc the potential equalization cable is to be dis...

Страница 93: ...h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 93 of 197 7 4 6 4 Plug assignment at the UserTerminal If the UserTerminal was disassembled dur...

Страница 94: ...5 v1_07hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 94 of 197 De...

Страница 95: ...5 v1_07hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 95 of 197 De...

Страница 96: ...0130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 96 of 197 For devices with...

Страница 97: ...ated 07 02 2013 printed 04 03 2013 page 97 of 197 MCU5 control board and display board inside the UserTerminal Sticker for plug references at the back plate of the UserTerminal Sticker for plug refere...

Страница 98: ...p cosmos com created 07 02 2013 printed 04 03 2013 page 98 of 197 7 4 7 Cable connections models 200 300 75 125 venus saturn cos100710 socket in UserTerminal for inverter drive and elevation motor cos...

Страница 99: ...Pay attention to safety standard of host devices Read manual Always use potential isolation Do not connect medical devices with sports devices DE Gef hrliche Netzspannung Vor dem ffnen des Ger tes Net...

Страница 100: ...spective IEC standards e g IEC 950 for data processing equipment and IEC 60601 1 for medical equipment Furthermore all configurations shall comply with the valid version of the system standard IEC 606...

Страница 101: ...ires the manufacturer to inform the responsible organization which means the applier hospital medical practice etc about the following risks in the technical description N The connection of a PEMS to...

Страница 102: ...e integration of a system and to ensure that remaining potential risks are identified during the operation of the system For the system integrator this means N he has to plan the integration of all ME...

Страница 103: ...predictable abuse can occur N Is the connection to the network or data network carried out in accordance with the intended use or purpose according to 3 para 10 of the Medical Devices Act N Could a f...

Страница 104: ...l processes and applications for which a malfunction or interruption would lead to a critical situation for the patient as in an intensive care network are categorized as class C Such a network should...

Страница 105: ...citly in DIN EN 60601 1 3rd edition but the chapter on network and data network includes an approach to handling this development The standard requires that when using a network or data network with t...

Страница 106: ...ompleted the instruction protocol has to be signed by the h p cosmos technician and all trained persons and returned together with the signed delivery note and the registration form to h p cosmos No I...

Страница 107: ...he potential equalization cable has to be connected AFTER these measurements 06 Lubricate the running machine with 30 ml special silicone oil which is part of the service box 07 Check the tension of t...

Страница 108: ...Terminal For these models an additional keyboard or an external UserTerminal for remote control via RS232 is available as an option see chapter entitled Accessories For maintenance and diagnosis we re...

Страница 109: ...For all required measurements and control steps consider also local requirements of your country and see detailed instruction and protocol form order no cos11690xx 8 3 1 Visual inspections N Carry out...

Страница 110: ...e connectors have been damaged In this case the cable should be replaced and the device repaired immediately 8 3 3 Isolation resistance RISO measurement N Resistance between bridged L N and Protective...

Страница 111: ...tric safety tester based on IEC 60601 1 For all measurements read operation manuals of the measurement devices carefully and verify the values and intervals with the local guidelines and laws An appro...

Страница 112: ...y circuit It should be pointed out that these are not regulations standards or safety checks specifically for h p cosmos running machines but rather apply to all electrical devices and building instal...

Страница 113: ...the amount of 3 syringes each filled with 10 ml of silicone oil Special devices may have different oiling amounts see delivery papers In addition 1 syringe filled with air should be added in order to...

Страница 114: ...oil absorbed by measuring the difference in weight However the user should realize that this test can only show approximate values and trends and is not meant to replace the direct inspection by an a...

Страница 115: ...mp Devices without a UserTerminal are equipped with an automatic acoustic signal bleep indicating the device needs to be lubricated At the standard setting the acoustic signal will occur after every 1...

Страница 116: ...chines designed for cycling and wheelchair applications For those models a lubrication test tissue test according the instructions available from h p cosmos should be performed during the obligatory y...

Страница 117: ...ternal control unit indicates that the oil tank is nearly empty Refill the oil tank when the control lamp illuminates At standard settings the word OIL will flash after 1000 km of operation If this oc...

Страница 118: ...r if it has been adjusted wrongly In this case a backlash occurs between the driving shaft and the belt i e when weight is applied to the belt by stepping on it the belt slows down The belt tension sh...

Страница 119: ...shaft at the back of the device belt re entry zone Take care that long hair and loose clothing are not caught in the gap of the tail shaft N For safety reasons the adjustment procedure must be attend...

Страница 120: ...rol and tightening of the driving belt poly V belt system The correct tension of the driving belt is very important for proper operation of the running machine If the tension is too high this may caus...

Страница 121: ...ring oscillation N Check the tone in the display target 1G adjust the tension of the drive belt if necessary N Check the amplitude of the needle target 440 460 adjust the tension of the drive belt if...

Страница 122: ...ing disinfecting and before opening N Pay attention to the safety notes warnings and precautions of the running machine the accessories and the disinfectant The h p cosmos devices are neither sterile...

Страница 123: ...of the drive engine N devices manufactured since approx 2003 have hall effect sensors with metal toothed lock washer instead of light barriers N only for oversize running machines with a running surfa...

Страница 124: ...ction and focus A special small screwdriver is required for adjustment of the focus Adjustment direction and focus of the light barrier can be checked when the room is dark 8 13Spare parts consumables...

Страница 125: ...tling noises are heard during training please check if the machine has a firm stand and follow exactly the advice given in the chapter 7 4 5 Mechanical installation Most often false adjustment of the...

Страница 126: ...uches a metal piece keys or display there may be an electrostatic discharge between the user and the device Such discharge may interfere with the operation of the device Generally electrostatic discha...

Страница 127: ...flow will remain nearly unnoticed when bare metal parts for example screws are touched c If either sports or medical devices are defective e g because of a defective isolation inside the device the cu...

Страница 128: ...of the RS232 interface N Loop back test For testing the RS232 of the running machine a special RS232 test plug including testing instruction is available from the manufacturer Connect the plug to the...

Страница 129: ...UserTerminal Initiation Running belt is not moving One of the mode LED is flashing manual profile cardio test Message Description Explanation Display Activity E01 Oil interval due E 01 OIL flashing H...

Страница 130: ...nger than the high acceleration command SPRINT function or high acceleration level N Trouble shooting see administrator OPTION 48 N Contact authorized service engineer N Check voltage supply Do not us...

Страница 131: ...Running machine is still working N At MCU5 control electronic as of manufacturing date 12 2007 this error shows a defect of the option memory FRAM In this case please contact h p cosmos and request a...

Страница 132: ...Inverter Indication display UT Indication display FI Meaning of the error code related to the frequency inverter motor drive control function status 00 No alarm 10 E OC1 Overcurrent shut off during a...

Страница 133: ...act authorized service engineer Card Err 01 Error chip card drive not all models CARD OR S TART Running machine is still functioning except chip card function Chip card protocol active but no chip car...

Страница 134: ...et in the administrator level In this case please contact customer service Error messages at devices without UserTerminal Devices without UserTerminal are equipped with an automatic error message code...

Страница 135: ...in OP02 Selection of the OEM code By selecting the OEM code the functions and settings of the UserTerminal are adapted to the requirements of the listed OEM partner The following OEM codes can be sele...

Страница 136: ...y purposes e g for beginners or in case the ceiling height of the room s ceiling is insufficient the max elevation of every treadmill model can be limited indicates maximum elevation adjustable range...

Страница 137: ...ng belt movement it activates the motor break for 10 seconds Remark The motor brake works only on running machines with integrated speed measurement The speed sensor must be switched on OP11 1 If the...

Страница 138: ...erval for indication of required OIL service can be adjusted by means of OP 35 Standard 1000 km for treadmills 100 km for ladder ergometers OIL intervals are indicated with error code E01 OIL every 10...

Страница 139: ...the keys and has not yet been reached report current elevation angle recommended default setting OP40 Speed display as set point or true value 0 Display of speed as set point report target speed 1 Dis...

Страница 140: ...t of running machines types N Make sure that the correct device type is set at administrator OP03 N Go to administrator OP 05 and set maximum possible speed N Set the device in the manual mode to the...

Страница 141: ...er motor rotation or hall effect sensor with 60 incr per motor rotation elevation element with a 5 mm spindle for 0 25 elevation h p cosmos saturnsince 1994 3 elevation control increment disc with lig...

Страница 142: ...Y 2 digit number of year of production lot MM month of production lot NNNN 4 digit serial number OP98 Setting of limitation of usage time Secret option OP99 Test profile 99 for endurance testing in th...

Страница 143: ...nding on the device type in OP03 0 max elevation OP 08 Heart rate display interval 0 beat to beat 0 or 1 9 seconds OP 09 Display test for all displays OP 10 Enter the serial number of the device 300 1...

Страница 144: ...ion maintains 1 elevation gets down to 0 OP 46 Display and setting of elevation for ladder ergometer 75 degree 0 180 degree OP 47 Reset of oil and service interval OP 48 Running belt speed at maximum...

Страница 145: ...0 0 4165 30 0 0 119 0 200 xx 3 1 no limit 40 0 35 1 2617 40 0 3 120 0 180 3 2 50 0 1 2617 50 0 3 125 0 3 3 60 0 1 2617 60 0 3 100 0 3 4 80 0 1 2617 80 0 3 130 0 3 5 60 0 1 1700 60 0 3 100 0 3 6 80 0 1...

Страница 146: ...VA peak performances for milliseconds max 1s min during heel strike Drive motor capacity 3300 VA Elevation motor capacity not applicable 390 VA not applicable 390 VA 390 VA a not applicable 390 VA Sa...

Страница 147: ...Normally only nonconductive pollution Temporary conductivity caused by condensation is to be expected Central lubrication system manual Number of stored programmes 6 training profiles 10 test profile...

Страница 148: ...ax 1s min during heel strike Drive motor capacity 3300 VA Elevation motor capacity not applicable 500 VA not applicable 500 VA Safety standards c IEC EN 60335 1 VDE 0700 IEC EN 60601 1 2 IEC EN 60601...

Страница 149: ...be expected Central lubrication system manual Number of stored programmes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1...

Страница 150: ...input long time 2700 VA Power input momentary 5700 VA peak performances for milliseconds max 1s min during heel strike Drive motor capacity 4300 VA Elevation motor capacity 500 VA Safety standards n...

Страница 151: ...rogrammes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1 and COM2 standard COM3 optional standard for service COM4 n a u...

Страница 152: ...0 A 1 x 16 A Power input long time 7500 VA 41500 VA Power input momentary 15200 VA 66500 VA Drive motor capacity 11000 VA 30000 VA Elevation motor capacity 370 VA 18000 VA Safety standards n MDD direc...

Страница 153: ...700 1060 hPa Maximum operating altitude approx 10 000 feet 3000m without pressurization Pollution degree Degree 2 Normally only nonconductive pollution Temporary conductivity caused by condensation is...

Страница 154: ...least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be th...

Страница 155: ...t and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the mea...

Страница 156: ...via coscom v3 interface released for complete medical treadmill series ergospirometry system CORTEX CORTEX Biophysik GmbH Leipzig Germany ergo spirometry IIa CE0124 via coscom v3 interface released f...

Страница 157: ...e instead of the RS232 interface In that case an USB to RS232 interface adapter cable is available at h p cosmos under item no cos12769 To control via USB interface a processor Pentium 1 8 GHz or high...

Страница 158: ...in his work The arm supports with additional keyboard and additional stop button offers a wide range of applications e g exercising in gentle way for patients people with over weight orthopaedic probl...

Страница 159: ...ports is only admitted for walking and not for running N While running the arm supports have to be folded vertically downwards on the outside 1 The arm supports must never project into the area of the...

Страница 160: ...ndrail shortened at left side speed grip left cos14763 01 Handrail short cos11456 Handrail long rolled 2 pillars cos10166 Handrail detachable separable cos14192ral9007 Please quote the serial number o...

Страница 161: ...gly e g in a patient file We have deliberately refrained from offering standard settings for particular heights because the adjustments can widely differ according to the posture and the individual si...

Страница 162: ...inders and noted accordingly e g in a patient file We have deliberately refrained from offering standard settings for particular heights because the adjustment can widely differ according to the postu...

Страница 163: ...r 02 Push button Adjust the bar handrail to the desired height by pressing the push button 03 Move the bars horizontally to adjust the required width 04 Lock the locking lever by pushing it down 05 N...

Страница 164: ...under supervision N If the clamp handles are not locked sufficiently there is danger of squeezing due to unintentional movement of the therapist s seat or the foot support Open and overhanging clamp...

Страница 165: ...d supports the training itself N The unweighting can be levelled precisely to the needs of the patient and ensures an optimal training environment N High patient comfort for more motivation N The high...

Страница 166: ...f body motion They do not have straps undercutting during gait and have an optimal fitting due to the individually adjustable strap system Four different vests provide adaptation to all body sizes and...

Страница 167: ...ting the wheelchair on the treadmill and before using the wheelchair over ground or on the street Wheelchair stabilizer for deck length 200cm cos10227 250cm cos10226 300cm cos00096110031 Available for...

Страница 168: ...ldren Chest belt system S cos14903 02 S Chest belt system M cos14903 02 M Chest belt system L cos14903 02 L Chest belt system XL cos14903 02 XL to be replaced every 6 months or earlier at the first si...

Страница 169: ...e at the next start Change from forward to reverse rotation N Turn the key operated switch and start the machine with no one on the running belt and only when the belt is not moving N Observe the belt...

Страница 170: ...ll cord safety stop actuator or with the safety arch with chest belt system The length of the pull cord should be adjusted so that the subject cannot use the rear 30 of the running deck It limits the...

Страница 171: ...ng external power supply 90 260 VAC to 12 VDC 5 A cos10002 02 UserTerminal TouchPanel remote control for table with foot for desktop with PC software h p cosmos para control connected via 5 meter cabl...

Страница 172: ...cy stops If there is no socket for external emergency stops order the special rear panel terminal for additional keyboard cos10109 cos15294L10m incl 10 m cable no spiral cable please stipulate the ser...

Страница 173: ...that case a USB to RS232 interface adapter cable is available at h p cosmos under order number cos12769 To control via USB interface the processor must be a Pentium 1 8 GHz or higher Interface cable...

Страница 174: ...ransmission Precision of heart rate measuring 1 resp 1 beat per minute Pay attention to the advice regarding safety and warnings about possible interferences in the chapters on errors technical data E...

Страница 175: ...you should keep a gap of at least 100 cm between the devices to exclude interference Transmission problems N The setting POLAR logo upside down is recommended for approximately 10 15 of the subjects N...

Страница 176: ...ured STOP and QUICK STOP button on the UserTerminal ARE NO EMERGENCY BUTTONS which are disconnecting from the mains power supply They are no emergency control based on IEC 60601 1 2005 Most of the add...

Страница 177: ...urance parameters online The treadmill can be controlled automatically The software is not designed for medical diagnostics or evaluation The treadmill data can support the doctor s decision Data can...

Страница 178: ...e clear design of the program allows an easy evaluation The comprehensive printout simplifies the interpretation and is the basis for best possible coaching The software is not designed for medical us...

Страница 179: ...f the most common gait and motion specific disorders with the possibility to transfer to the individual report N Integrated player for the analysis of several video clips N Video overlay of up to eigh...

Страница 180: ...priate tools eye protection dust mask etc Contact service h p cosmos com to receive the safety data sheet of the European Commission Directive 91 155 EEC h p cosmos running machine h p cosmos running...

Страница 181: ...admill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 181 of 197 14 Appendix 1 Certificates Latest versions always http www h...

Страница 182: ...en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 182 of 197 14 2Certific...

Страница 183: ...5 v1_07hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 183 of 197 1...

Страница 184: ...est 17 34 1054 seconds Heart rate at 500 m 158 bpm 1000 m 156 bpm 1500 m 160 bpm and 2000 m 155 bpm Average heart rate 158 156 160 155 4 157 bpm Fitness Index 420 50 x 0 2 1054 x 0 19338 157 x 0 56 10...

Страница 185: ...y note and the registration form are to be sent to h p cosmos Instruction on general operation 1 Hand over operating manual Inform about operating and service manual always keep one within reach of al...

Страница 186: ...N OFF button at the UserTerminal 5 Explain safety features that the delivered model is equipped with for example emergency stop quick stop safety harness etc Do not use the emergency stop button for n...

Страница 187: ...mode and the integrated test programs inform about interface and compatibility list Potential isolation for any interface linkage is obligatory in the field of medicine Only connect to coscom v3 compa...

Страница 188: ...ect adjustment of levelling sockets Check weight load on the levelling sockets by trying to lift up the treadmill frame 15 Proper adjustment of running belt with regulation screw and allen key For eac...

Страница 189: ...e performed by authorised and trained personnel who are certified by h p cosmos 19 Inform the user about periodically recurring maintenance intervals and obligatory safety controls as well as the dang...

Страница 190: ...tended and forbidden use safety notes or precautions as well as unauthorized maintenance or lack thereof and or regular safety checks may lead to injuries or even death and or can damage the device Th...

Страница 191: ...10m5 v1_07hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 191 of 19...

Страница 192: ...5 v1_07hpc en 20130207_cos14310m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 192 of 197 1...

Страница 193: ...0m5 v1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 193 of 197 15 6Wiring diagram load circuit...

Страница 194: ...1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 194 of 197 15 7Wiring diagram control circuit q...

Страница 195: ...1 07hpc en_h p cosmos_treadmill_manual doc 2013 h p cosmos sports medical gmbh author fh email h p cosmos com created 07 02 2013 printed 04 03 2013 page 195 of 197 15 8Wiring diagram control circuit q...

Страница 196: ...afety information N Earth IEC 60417 5017 Warning safety precautions Pay attention to accompanying instructions ISO 7010 W001 Non sterile product 2012 Manufacturer and year of manufacture P Potential c...

Страница 197: ...ally Skype with webcam Service department phone 49 86 69 86 42 0 phone direct 49 86 69 86 42 25 mobile 49 171 720 69 88 fax 49 86 69 86 42 49 email service h p cosmos com skype h p cosmos com search s...

Отзывы: