EXTERNAL CONTROL
117
Inquires about the set output power of a single channel
model.
:POW:A?
Inquires about the set output power of CH1.
:POW:B?
Inquires about the set output power of CH2.
Response
example
100
Indicates that the set output power is 100W.
Remark
When the tracking mode is on, setting CH1 will affect
CH2. Setting CH2 in the tracking mode will cause an
execution error.
OUTPUT ON/OFF (
:
OUTP)
This command turns on or off output.
Setting
:OUTP <value>
Sets the output of a single channel model.
:OUTP:A <value>
Sets the output of CH1.
:OUTP:B <value>
Sets the output of CH2.
The relationship between <value> and On/Off is as
shown below:
1: Output On, 0: Output Off
On the dual channel model, :OUTP selects the OUTPUT.
The output is activated only after <value> is set to 1.
Application
example
:OUTP 1
Activates output.
Query
:OUTP?
Inquires about the output setting of a single channel
model.
:OUTP:A?
Inquires about the output setting of CH1.
:OUTP:B?
Inquires about the output setting of CH2.
Response
example
1
Indicates that the output is activated. The return values
of the query have the same meaning as the setting
<value>.
Remark
This command causes an execution error if the output is
externally controlled.
Sending the :OUTP command causes a command error
when both channels of the dual channel model are
deactivated.