
OPERATION
137
Select CV Response Speed
Background
Constant voltage mode has fast and slow response
speeds. Quick current changes can induce line
voltage, making it more difficult for the PEL load
generator to maintain a constant current. In these
types of conditions, slow response speed is
recommended.
Maximum current range is dependent on the load
module type.
Ensure the menu is in CV Mode. See page 134.
Panel Operation
1.
Press F3 (Response) to switch
between Fast and Slow response
speeds.
Response speed settings will be
reflected in the Current Operation
Channel Status panel.
Slow response
speed
Slow
Fast response
speed
Fast
Note
Changing the response speed will only affect the
current (active) channel. Other channels will not be
affected by any changes.