GSP-830 User Manual
92
2b Deactivating
limit line table
Press F2 (Edit Table Off). The
editing table disappears and the
edited contents will be saved.
Edit Table
On Off
F 2
3a. Add a limit line
point
1.
Make sure that the cursor is
pointing to the first empty
frequency point. Use the
Up/Down keys to move the
cursor, if necessary.
10 points are available for each
high and low limit line.
No.
MHz
dBm
1
2
3
4
5
2.
If necessary, move the cursor to
different frequency points using
the Arrow keys.
3.
Enter the frequency using the
Numerical and Unit keys.
Range: 9kHz to 3GHz
The value is displayed in MHz.
BK SP
dB
Enter
GHz /
Sec
mSec
MHz /
kHz /
Sec
μ
4.
The cursor automatically moves
to the dBm side. Enter the gain
in using the Numerical keys and
dB or Enter key.
Range:
−
130dBm to +20dBm
No.
MHz
dBm
1
2
3
4
5
98
5.
Continue the above steps for
the number of points required.
3b. Delete a limit
line point
1.
Move the cursor using the
Arrow keys.
2.
Press F3 (Delete) to delete the
point (frequency and amplitude
together).
Delete
F 3