Adjustment Procedure (Manual procedure) 

Model no. 

GDM

–825XA 

Document No.  

TO 

– XX -XXX 

Page 1 of  

NO. 

Item 

Condition 

 

Appendix 3 

2WR 

Connection 

a. 2WR is a connection method for 2 wires. 

b. When 2WR is performed, EX SNS in Fluke5720A should be turned 

OFF

  

c. 2WR connection method is shown as the diagram below. 

 

Appendix 4 

Common 

problems 

1. If there is no display on the screen after turning the power on 

Please 

plug out the Calibration Key, and turn the power on again. 

2. If the resistance value is unstable at the 

100MΩreange 

 When 

calibrating the high resistance, users need to use a high-voltage 

insulated cable (over 1KV rating) as the test connection cable. 

3. If the device is still not accurate after all resistance ranges have been 

adjusted 

Because the 0R and OPEN affect all resistance ranges 

parameter, users need to adjust CAL CODE1 through to 9 all at once. 

You cannot calibrate each CAL CODE 1 separately. 

4. AC10mV & AC1mA is not accurate 

 Because AC10mV and AC1mA 

is calculated from the ATE program, users can

’t adjust this manually. If 

AC10mV or AC1mA is not accurate, it is necessary to use the ATE 

program to calibrate them again. 

 