Adjustment Procedure (Manual procedure) 

Model no. 

GDM

–825XA 

Document No.  

TO 

– XX -XXX 

Page 1 of  

NO. 

Item 

Condition 

 

  

Step 4 

END 

1.  After adjustment is complete, take out the Calibration Key 

(GDM-01) from the CAL Key Port and reset the power switch.  

 

Appendix 1 CAL CODE 

CAL CODE 

Connection/
Function 

8255A 

8251A  Remark 

Please refer to appendix 2 for 4WR connection

 

4WR 

0R 

0R 

It is necessary to use 4 wire 
resistance measurement to 
compensate for EMF. Wait for 60 
seconds before starting adjustment

 

4WR 

100R 

100R 

 

 

4WR 

1KR 

1KR 

 

 

4WR 

10KR 

10KR 

 

 

4WR 

100KR 

100KR 

 

 

4WR 

1MR 

1MR 

 

 

4WR 

10MR 

10MR 

 

 

Please refer to appendix 3 for 2WR connection

 

2WR 

2W100MR 100MR 

 

 

Remove all inputs from the terminals 

 

OPEN 

OPEN 

OPEN 

 

 

Use the INPUT V and the COM terminal for DCV or ACV calibration. 

 

11 

DCV 

0V 

0V 

 

 

12 

DCV 

190mV 

100mV 

 

 

13 

DCV 

-190mV 

-100mV 

 

 

14 

DCV 

1.9V 

1V 

 

 

15 

DCV 

19V 

10V 

 

 

16 

DCV 

190V 

100V 

 

 

17 

DCV 

750V 

750V 

 

 

61 

ACV 

19mV 

19mV 

1kHz 

62 

ACV 

190mV 

100mV  1kHz 

63 

ACV 

1.9V 

1V 

1kHz 

64 

ACV 

19V 

10V 

1kHz 

65 

ACV 

190V 

100V 

1kHz 