ASR-3000 Series User Manual
72
When it is under AC-INT, AC-ADD or AC-Sync
mode.
Use the scroll wheel or the numerical keypad to set
value of Vrms limit directly or use the F3 (MAX)
and F4 (MIN) soft-keys to set the limit to the
maximum or minimum value.
AC–INT, AC-ADD, AC-Sync
Vrms
Range
10% ~ 100% full range
voltage
Soft-keys
MAX, MIN
Vrms
Setting
Note
The Vrms Limit value defined by user will be generally
applied to AC-INT, AC-ADD and AC-Sync modes under
the same voltage range, which divides into 2 levels,
high range including AUTO and 200V and low range
covering 100V.
3.
Set the voltage limit (VPK+ & VPK-) with the scroll
wheel or the numerical keypad or with the F3
(MAX) and F4 (MIN) soft-keys to set the limit to
the maximum and minimum values, respectively.
AC+DC-INT, DC-INT,
AC+DC-ADD, AC+DC-Sync
VPK+
Range
3.5% ~ 100% full range peak
voltage
Soft-keys
MAX, MIN
VPK-
Range
3.5% ~ 100% full range peak
voltage