SEPARATION IN THE CONTRACTION CHAMBER

.5

.5

Separation  of  mercury  in  the  contraction  chamber  is  an  entirely  different  problem.    Many 

thermometers  contain  scale  ranges  which  begin  well  above  ambient  temperatures.    These 

thermometers  are  provided  with  a  contraction  chamber  or  enlargement  which  prevents  the 

mercury from entering the bulb at ambient temperatures.  Unusual handling of the themometer 

may cause a separation in the contraction chamber.  

 

The procedure to rejoin this mercury is relatively simple.  If the separated mercury is in the 

form of a very slight amount, invert the thermometer and tap gently against the palm of the hand.  

(It  is  recommended  that  protection  be  used  in  the  palm  of  the  hand  -  paper  towel,  cloth,  etc. 

against glass breakage.)

  This will cause a greater separation, adding volume and weight to the 

separated portion.  The thermometer should then be righted and gripped firmly in the hand with 

bulb outward.  Sharply swing the thermometer downward at one's side in a 270 to 180 degree arc.  

The centrifugal force thus generated will join the separated mercury to the main portion.  If the 

original  separation  contained  a  large  volume,  only  the  latter  step  of  swinging  the  thermometer 

downward need be followed. 

 

SEPARATION IN THE LOWER AND MIDDLE PORTION OF THE COLUMN

.5

.5

Separations of this type are less frequent and more difficult to repair.  There are more variables, 

and no one explanation will cover all types of thermometers.   

 

The general procedure is to subject the bulb only to a temperature (in an ice bath or freezer) 

sufficient  to  retract  all  the  mercury  into  the  bulb.    A  slow  and  careful  return  to  ambient 

temperature will return an intact column.  Our experience has shown that additional problems may 

be caused in these procedures, and we point out the following cautions: 

 

Allow  thermometer  to  return  to  ambient  or  room  air  slowly,  or  breakage  may  occur.    This 

breakage is caused by mercury thawing in the capillary more slowly than the mercury in the bulb 

thereby creating an impasse to the expanding mercury.  In order to avoid this breakage, extreme 

care must be taken to allow the mercury in the bulb to liquify at the same rate as the mercury in the 

capillary. 

 

Position  the  thermometer  vertically  and  allow  it  to return to ambient temperatures.  Extreme 

care  must  be  taken  to  insure  the  thermometer  is  not  jarred  or  at  the  slightest  angle  while  the 

nitrogen gas and the mercury return into the capillary.  If this care is not taken, mercury will enter 

the capillary prior to the gas.  This will cause gas bubbles to develop in the bulb and, in turn, cause 

considerable inaccuracy which may be undetectable when the thermometer is placed into use. 

 

1)  

If there is no separation, return the unit to Guth Laboratories, 

Inc. (See instructions page 9). 

 

2)  

If  there  is  a  separation...Try  another NIST thermometer (or) 

follow the instructions for rejoining a mercury separation. 

 

CAUTION

: If a mercury  thermometer  is  accidentally  broken,    be  very  careful.    Mercury  is  

poisonous..  

DO NOT

allow direct contact with skin or mouth!  Use rubber gloves to clean any 

spill. 

 

INSTRUCTIONS  FOR  REJOINING  MERCURY  SEPARATIONS  IN  GLASS 

THERMOMETERS 

 

All  Guth  precision  thermometers  are  sealed  under  extremely  high  nitrogen  pressure  while 

drawing the mercury into the lower portion of the capillary.  This process, when coupled with acid 

washed  tubing  and  triple  distilled  mercury,  although  costly,  greatly  reduces  the  frequency  of 

separations in the capillary portion of the thermometer.   

 

Many instructions for rejoining separations begin with "the mercury may be retracted into the 

bulb  by  immersing  it  in  a  suitable  mixture  of  dry  ice  and  alcohol..."    This  statement  is  NOT 

specific  enough.    Determine  the  type  of  separation  from  the  illustrations  provided  and  proceed 

according to these instructions.  

 

SEPARATION AT THE UPPER PORTION OF MERCURY COLUMN

.5

.5

All well constructed thermometers contain an expansion chamber or bubble at the extreme top 

of the capillary.  This chamber serves a twofold purpose; one, to accommodate an overflow of 

mercury when the thermometer is subjected to temperatures in excess of its scale range and two, as 

a means of rejoining this type of separation. 

 

With  the thermometer in a vertical position, slowly heat the bulb under WARM tap water until 

the separated segments and a portion of the main (intact) column enter the top chamber. (Extreme 

care  must  be  taken  to  insure  that  the  mercury  does  NOT  fill  more  than  HALF  to  THREE-

QUARTERS of the chamber volume; 

otherwise, breakage of the bulb will result

).

 

The nitrogen 

pressure  will  force  a  rejoining  of  the  mercury.    While  positioning  the  thermometer  vertically, 

examine the column as it cools and retracts to be sure it is intact. 

 

7

8

Содержание 2100

Страница 1: ...Guth Laboratories Inc MODEL 2100 SIMULATOR OPERATION MANUAL 590 North 67th Street Harrisburg PA 17111 4511 TELEPHONE 717 564 5470 800 233 2338 FAX 717 564 2555 WEB SITE AT www guthlabs com PRICE 5 00 ...

Страница 2: ... 1999 Guth Laboratories Inc all rights reserved ...

Страница 3: ...ance To establish a precise and accurate calibration standard when using a Guth simulator GUTH LABORATORIES INC recommends using Guth Certified Simulator Solution Guth Laboratories Inc a pioneer and leader in the science of alcohol breath testing has provided Certified Simulator Solution to state municipal and local law enforcement agencies for many years Guth Premixed Certified Simulator Solution...

Страница 4: ...ature and heater activity x RELIABLE AGITATOR MOTOR x AGITATOR PADDLE Designed by Guth Laboratories Inc Offers maximum circulation and uniformity of solution x HEATER ELEMENT Electronically protected from overheating when removed from solution or when temperature sensor is defective x RFI PROTECTED The heating element and motor will turn off when the simulator is subjected to a RF field of 10 volt...

Страница 5: ...MODEL 21 00 ON OFF RESET GUTH LABORATORIES INC 5 5 GUTH MODEL 21 00 SIMULATOR C 3 ...

Страница 6: ...MODEL 21 00 REAR VIEW CAUTION REMOVE SOLUTION BEFORE MOVING INSTRUMENT REPLACE FUSE WITH BUSS TYPE GMW 1amp 125 vac Fast Acting 4 ...

Страница 7: ...ompletely illuminate as shown in figure 1 This indicates the display is functioning properly Figure 1 6 The L E D test display in figure 1 will remain for approximately one second and then change to the COLD display as shown in figure 2 Figure 2 7 Scanning of the decimal dots across the bottom of the display indicates the simulator is heating 8 When the solution reaches 33 6o C the display will in...

Страница 8: ...a simulator bellows pump a pump powered air source or an atomizer Several models of pumps are available through Guth Laboratories If a low reading is obtained check all tubing for leaks Be sure the glass container is properly tightened to the metal top housing Loosen and re tighten the laboratory glass container DO NOT OVERTIGHTEN Repeat the testing procedure IF THE SIMULATOR DOES NOT HEAT TO 34o ...

Страница 9: ...rsing it in a suitable mixture of dry ice and alcohol This statement is NOT specific enough Determine the type of separation from the illustrations provided and proceed according to these instructions SEPARATION AT THE UPPER PORTION OF MERCURY COLUMN 5 5 All well constructed thermometers contain an expansion chamber or bubble at the extreme top of the capillary This chamber serves a twofold purpos...

Страница 10: ...ON IN THE LOWER AND MIDDLE PORTION OF THE COLUMN 5 5 Separations of this type are less frequent and more difficult to repair There are more variables and no one explanation will cover all types of thermometers The general procedure is to subject the bulb only to a temperature in an ice bath or freezer sufficient to retract all the mercury into the bulb A slow and careful return to ambient temperat...

Страница 11: ... PA 17111 4511 717 564 5470 or 800 233 2338 NOTE When returning the simulator it is imperative the laboratory glass container is attached This is necessary to protect critical parts of the simulator Remove the glass container empty contents dry thoroughly and replace container Carefully wrap the simulator in bubble wrap newspaper paper towels etc Package the simulator well placing packing material...

Страница 12: ...play the error code This will continue until the malfunction is corrected or the simulator is reset RESETTING THE SIMULATOR Turning the power off and on will reset the simulator ERROR CODES The following list of error codes indicate the cause and corrective action REMEDY Turn the simulator off Fill container with 500 ml of certified solution and reassemble top housing and container REMEDY Cannot b...

Страница 13: ...wait 3 5 seconds then turn on REMEDY Reset the electronics Turn off for 3 5 seconds then turn on If the simulator does not heat the heating element may be defective Requires service The temperature of the solution is above 34 2qC or radio frequency interference RFI has been detected Electronics need to be reset Excessive amount of time to obtain operating temperature longer than 15 minutes The hea...

Страница 14: ...eating element may be defective Requires service REQUIRES SERVICE Simulator must be repaired by a Guth certified technician or returned to Guth Laboratories Inc Seeinstructions on page 9 The temperature of the solution is below 33 8o C after it has initially obtained 34 0o C 1 12 ...

Страница 15: ... IN MOTOR HE ATE R MOTOR C ONT ROL HE ATE R C ONT ROL BUFFER S TR ANSFORMER POWE R SUPPL Y 5vdc TO ALL CIR CUITS I C BUS 2 A to D CONVE RTE R VOL T AGE R E FE R ENCE 4 096vdc 3 WIREBUS CALIBR ATION TEST POINT RFI DETE CT HEATE R THE R MOCOUPLE HIGH T E MPE RATURE SHUT OFF AMPLIFIE R C OMPAR ATOR BE E PER LED INSTR UMENT AMPLIFIER OUTPUT SE CT ION INPUT SE CTION PR EC ISION THER MISTOR 13 ...

Страница 16: ...__ CERTIFIED BY________________________________DATE_____________ TEST TEMPERATURE_______qC NIST TRUE TEMPERATURE_______qC MODEL 2100 SERIAL NUMBER_________________________________ CALIBRATION CHECK DATE CHECKED BY NOTE Only a Guth Laboratories Inc certified technician shall perform Field calibration 14 ...

Страница 17: ...hanging solution Disconnect power cord before replacing fuse Fuse Replacement Buss type GMW 1 amp 125 vac Fast Acting Indoor Use Only Altitude up to 2000m 6562 ft Temperature 5 C to 33 C 41 F to 91 F Maximum Relative Humidity 80 for Temperatures up to 31 C 88 F decreasing linearly to 50 Relative Humidity Transient Over voltages according to Installation Category II Pollution Degree 2 in accordance...

Отзывы: