RECORDS (SERIAL NUMBERS, CALIBRATION  AND NOTES)

NIST TRACEABLE THERMOMETER – SERIAL NUMBER___________________ 

 

CERTIFIED BY________________________________DATE_____________ 

 

TEST TEMPERATURE_______

q

C

NIST TRUE TEMPERATURE_______

q

C

MODEL 2100  –  SERIAL NUMBER_________________________________ 

 

CALIBRATION CHECK DATE

CHECKED BY

 

NOTE: Only a Guth Laboratories, Inc. certified technician shall perform Field 

calibration.

 

INTRODUCTION

The Guth Model 2100 Alcohol Breath Simulator is a State-of-the-Art, electronically temperature 

controlled, water-alcohol instrument for  the   purpose of providing a precise calibration standard.  

Temperature of the water-alcohol solution placed in the simulator is maintained at 34

o

C + .05

o

C

or better. 

 

The  Guth  Model  2100  Simulator  is  approved  by  the  National  Highway  Traffic  Safety 

Administration (NHTSA), published on their Conforming Products List (CPL), and is listed with 

Underwriters Laboratories, Inc. (UL), as meeting United States Safety Standards and Canadian 

Safety Standards (CUL). 

 

CERTIFIED*  SIMULATOR  SOLUTION  STANDARD

 

PLEASE NOTE: 

It is imperative your Simulator Solution be of the highest quality in 

order to acquire the scientific and legal standards of acceptance

.

To  establish  a  precise  and  accurate  calibration  standard  when  using  a  Guth  simulator,  GUTH 

LABORATORIES,  INC.  recommends  using  Guth  Certified  Simulator  Solution.  Guth 

Laboratories,  Inc.,  a  pioneer  and  leader  in  the  science  of  alcohol  breath  testing,  has  provided 

Certified  Simulator  Solution  to  state,  municipal,  and  local  law  enforcement  agencies  for  many 

years. 

 

* * * * * * * * * * * *

Guth Premixed Certified Simulator Solution is the only solution available which is prepared by 

a FDA / DEA Registered Laboratory. 

 

* * * * * * * * * * * *

Guth Certified Simulator Solutions are prepared in standard concentrations of .02%, .04%, .05%, 

.08%,  .10%,  .15%  and  .20%.    Other  special  concentrations  are  available.    Please  contact  Guth 

Laboratories, Inc. for availability and pricing: 

 

Toll free:  (800) 233-2338   

Fax:  (717) 564-2555 

 

or visit our web site at: www.guthlabs.com 

 

STORAGE:

 

Store Simulator Solution in a cool location.  Do not place solution in a freezer 

DO NOT REFRIGERATE SIMULATOR. 

 

*Guth Certified Simulator Solutions are tested by recognized laboratories. 

 

1

14

Содержание 2100

Страница 1: ...Guth Laboratories Inc MODEL 2100 SIMULATOR OPERATION MANUAL 590 North 67th Street Harrisburg PA 17111 4511 TELEPHONE 717 564 5470 800 233 2338 FAX 717 564 2555 WEB SITE AT www guthlabs com PRICE 5 00 ...

Страница 2: ... 1999 Guth Laboratories Inc all rights reserved ...

Страница 3: ...ance To establish a precise and accurate calibration standard when using a Guth simulator GUTH LABORATORIES INC recommends using Guth Certified Simulator Solution Guth Laboratories Inc a pioneer and leader in the science of alcohol breath testing has provided Certified Simulator Solution to state municipal and local law enforcement agencies for many years Guth Premixed Certified Simulator Solution...

Страница 4: ...ature and heater activity x RELIABLE AGITATOR MOTOR x AGITATOR PADDLE Designed by Guth Laboratories Inc Offers maximum circulation and uniformity of solution x HEATER ELEMENT Electronically protected from overheating when removed from solution or when temperature sensor is defective x RFI PROTECTED The heating element and motor will turn off when the simulator is subjected to a RF field of 10 volt...

Страница 5: ...MODEL 21 00 ON OFF RESET GUTH LABORATORIES INC 5 5 GUTH MODEL 21 00 SIMULATOR C 3 ...

Страница 6: ...MODEL 21 00 REAR VIEW CAUTION REMOVE SOLUTION BEFORE MOVING INSTRUMENT REPLACE FUSE WITH BUSS TYPE GMW 1amp 125 vac Fast Acting 4 ...

Страница 7: ...ompletely illuminate as shown in figure 1 This indicates the display is functioning properly Figure 1 6 The L E D test display in figure 1 will remain for approximately one second and then change to the COLD display as shown in figure 2 Figure 2 7 Scanning of the decimal dots across the bottom of the display indicates the simulator is heating 8 When the solution reaches 33 6o C the display will in...

Страница 8: ...a simulator bellows pump a pump powered air source or an atomizer Several models of pumps are available through Guth Laboratories If a low reading is obtained check all tubing for leaks Be sure the glass container is properly tightened to the metal top housing Loosen and re tighten the laboratory glass container DO NOT OVERTIGHTEN Repeat the testing procedure IF THE SIMULATOR DOES NOT HEAT TO 34o ...

Страница 9: ...rsing it in a suitable mixture of dry ice and alcohol This statement is NOT specific enough Determine the type of separation from the illustrations provided and proceed according to these instructions SEPARATION AT THE UPPER PORTION OF MERCURY COLUMN 5 5 All well constructed thermometers contain an expansion chamber or bubble at the extreme top of the capillary This chamber serves a twofold purpos...

Страница 10: ...ON IN THE LOWER AND MIDDLE PORTION OF THE COLUMN 5 5 Separations of this type are less frequent and more difficult to repair There are more variables and no one explanation will cover all types of thermometers The general procedure is to subject the bulb only to a temperature in an ice bath or freezer sufficient to retract all the mercury into the bulb A slow and careful return to ambient temperat...

Страница 11: ... PA 17111 4511 717 564 5470 or 800 233 2338 NOTE When returning the simulator it is imperative the laboratory glass container is attached This is necessary to protect critical parts of the simulator Remove the glass container empty contents dry thoroughly and replace container Carefully wrap the simulator in bubble wrap newspaper paper towels etc Package the simulator well placing packing material...

Страница 12: ...play the error code This will continue until the malfunction is corrected or the simulator is reset RESETTING THE SIMULATOR Turning the power off and on will reset the simulator ERROR CODES The following list of error codes indicate the cause and corrective action REMEDY Turn the simulator off Fill container with 500 ml of certified solution and reassemble top housing and container REMEDY Cannot b...

Страница 13: ...wait 3 5 seconds then turn on REMEDY Reset the electronics Turn off for 3 5 seconds then turn on If the simulator does not heat the heating element may be defective Requires service The temperature of the solution is above 34 2qC or radio frequency interference RFI has been detected Electronics need to be reset Excessive amount of time to obtain operating temperature longer than 15 minutes The hea...

Страница 14: ...eating element may be defective Requires service REQUIRES SERVICE Simulator must be repaired by a Guth certified technician or returned to Guth Laboratories Inc Seeinstructions on page 9 The temperature of the solution is below 33 8o C after it has initially obtained 34 0o C 1 12 ...

Страница 15: ... IN MOTOR HE ATE R MOTOR C ONT ROL HE ATE R C ONT ROL BUFFER S TR ANSFORMER POWE R SUPPL Y 5vdc TO ALL CIR CUITS I C BUS 2 A to D CONVE RTE R VOL T AGE R E FE R ENCE 4 096vdc 3 WIREBUS CALIBR ATION TEST POINT RFI DETE CT HEATE R THE R MOCOUPLE HIGH T E MPE RATURE SHUT OFF AMPLIFIE R C OMPAR ATOR BE E PER LED INSTR UMENT AMPLIFIER OUTPUT SE CT ION INPUT SE CTION PR EC ISION THER MISTOR 13 ...

Страница 16: ...__ CERTIFIED BY________________________________DATE_____________ TEST TEMPERATURE_______qC NIST TRUE TEMPERATURE_______qC MODEL 2100 SERIAL NUMBER_________________________________ CALIBRATION CHECK DATE CHECKED BY NOTE Only a Guth Laboratories Inc certified technician shall perform Field calibration 14 ...

Страница 17: ...hanging solution Disconnect power cord before replacing fuse Fuse Replacement Buss type GMW 1 amp 125 vac Fast Acting Indoor Use Only Altitude up to 2000m 6562 ft Temperature 5 C to 33 C 41 F to 91 F Maximum Relative Humidity 80 for Temperatures up to 31 C 88 F decreasing linearly to 50 Relative Humidity Transient Over voltages according to Installation Category II Pollution Degree 2 in accordance...

Отзывы: