Set your emulator to 8 data bits, 1 stop bit and no parity bits (8-N-1) and
disable both hardware (XON/XOFF) and software (RTS/CTS) flow control.
2.3.3.2 Data
The DATA connector is referred to throughout the Platinum manual as “DATA
OUT”. It is a general purpose serial port and is fully configurable within
Platinum as PortDO.
2.3.3.3 USB
The USB connectors can be used to connect up to seven CD24-R8 digitisers
and at most one external storage device, such as a USB thumb-drive.
CD24-R8 digitisers connected to these ports are automatically configured as
GCF input sources. Entries will appear for each on the System Status page of
the web interface and appropriate entries will appear in the instrument
control and digitiser configuration menus. If the digitisers are connected and
powered-up after the EAM-R has been started, you will have to refresh the
web page in order to see these entries.
Note:
In order for this feature to work correctly, the DATA
OUT ports of the CD24-R8 digitisers must be configured to
operate with a line speed of 115,200.
A USB storage device connected to any of the USB ports becomes available to
Platinum as “External USB storage”. It can be selected as the recording device
using the “Configuration”
“Storage and recording” page of the web
→
interface. Please see the “Recording and Retrieving Data” chapter of the
Platinum manual for more details.