Operator's Guide
retrieve control information and display the
Control
window.
The options you can control immediately are:
•
the type of sensor you are using,
•
GPS power cycling options,
•
the short-term and long-term average values for
triggering (but
not
which streams perform the trigger, or
which are output by it)
•
the length of pre-trigger and post-trigger periods,
•
calibration signal options, and
•
mass control functions.
Some of these options can also be altered in the
Configuration
setup
window.
For more information on how to use Scream! to configure your
digitizer, please see the Scream! user guide or the on-line help.
If you need a more powerful interface to the DM16, you can
also issue commands to it directly using Scream!'s terminal
mode. A terminal window is opened by right-clicking on the
digitizer's entry in the list and selecting
Terminal...
.
The digitizer will stop transmitting data while you have a
terminal window open. Data collected whilst the terminal is
active will be stored for transmission later, memory space
permitting.
5.2 Terminal commands
You can connect to the internal software of the DM16 over its
output serial port and communicate with it.
To enter command mode from Scream!, right-click on the
digitizer's icon and select
Terminal...
from the menu that pops
up. A window will open, and once the digitizer and computer
are communicating properly you will see the prompt
ok
If you prefer, you can use a terminal program on your computer
(such as
minicom
on Linux, or
hypertrm
on Microsoft Windows)
July 2016
27