5 .04 Troubleshooting
If the LP11 - Side Exit does not react to use of the operation panel
• Check that the emergency stop has not been activated. The emergency stop
is released by turning the knob until it pops out
• Check that the battery box is charged and connected
• Contact the Guldmann Service Team or the distributor
If the LP11 - Side Exit can go up, but not down
• Remove impurities from the safety jam strip
• Check the safety jam strip for other blockages
If the LP11 - Side Exit starts to beep
• An audio signal is a sign of critically low battery levels. The batteries must
therefore be charged before the platform is used again
If the LP11 - Side Exit stops and starts beeping
• If the platform is overloaded, it stops and starts beeping.
• Remove some of the load from the platform
• Observe the maximum load indication of the platform