© Guldmann
GB/US
-09/2016 •
#550376_3
6
GENERAL INFORMA
TION
General conditions
Product liability
In relation to product liability the current rules under Danish
law shall apply .
The Seller shall never be held liable for operating losses,
loss of earnings and other indirect losses .
Product liability insurance which includes cover against
injury and property damage has been taken out .
Any disputes between the parties shall be settled by the
Maritime and Commercial Court in Copenhagen (
Sø- og
Handelsretten
) in pursuance of Danish law
Transport and storage
Transport and storage must be in the range of limited
environmental condition:
Temperature between -10 and +40°C / 14 and 104°F
Relative air humidity between 30 and 70 %
Air pressure between 70 and 106 kPa
(See symbols page 8)
Warning!
Do not modify this equipment without authorization of the
manufacturer .
Installation in corrosive surroundings
Special guidelines must be observed when installing the
rail system around swimming pools, etc .
The surfaces of the brackets used, traverse travelling
trolleys, etc ., must be painted . Hoists must be supplied
with a special surface treatment .
All fixing elements, nuts, washers, threaded rods, must be
acid-proof .
Acid-proof threaded rods and dual-component adhesive
must be used for installation in concrete, tiles, etc .
Through-going threaded rods or acid-proof screws must
be used for installation in wood .
Particular care should be taken to avoid damaging the
surface of painted components .
Washers made of synthetic material must be used in
between painted surfaces and nuts and ordinary washers .
Damaged paint must be repaired after installation, and
visible threaded rods, nuts, washers, etc ., must be covered
with paint .
Hidden threaded rods, nuts, washers, and brackets in the
rail track must be treated with a silicone coating .
Wherever possible, nuts must be fitted with protective
caps .
There must be no swarf, dirt, or other impurities in the
rail system as they will accumulate moisture and lead to
corrosion .
Bearings in traverse travelling trolleys must be treated with
a silicone coating before installation .
Guldmann recommends that in corrosive surroundings like
swimming pools, bathing environments etc ., regular safety
and service inspection at rail systems and ceiling hoists
must be performed every ½ year . A complete overhaul of
the rail system (replacements of brackets, fixings etc .) can
be necessary approx . every 5 year .
Guldmann products are designed for use in these environ-
ments with the following maximum values of;
Water quality:
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7,6
Free chlorine, mg/l . . . . . . . . . . . . . . . . . . . . . . . . . . 2,0
Bound chlorine, mg/l . . . . . . . . . . . . . . . . . . . . . . . . 0,5
Air humidity:
The relative air humidity in the room should be approx .
70% at an outdoor temperature of approx . 20 degree C
and approx . 50% at an outdoor temperature of approx . . 0
degree C .
Note!
Quality designation for bolts, etc .
A4/AISI 316 = acid-proof
A2/AISI 304/rustproof 18 .8 = not acid-proof
Torque settings
All bolts and nuts are tightened with torque in accordance
with applicable rules – M8 = 27 Nm, M10 = 54Nm .
Hilti HUS-H 8 x 65 screws are tightened with a torque of
35 Nm .
Hilti HUS-H 8 x 85 – 150 screws are tightened with a torque
of 45 Nm .
The two M10 nuts at the GH Plasterboard Wall bracket are
thightened with a torque of 30Nm .
Содержание GH1 F
Страница 20: ... Guldmann GB US 09 2016 550376_3 22 GENERAL INFORMATION ...
Страница 140: ...WiFi Guldmann GB US 09 2016 550376_3 142 TECHNICAL DRAWINGS Wiring Diagram GH3 200 kg 250 kg drive WiFi ...
Страница 142: ...WiFi Guldmann GB US 09 2016 550376_3 144 TECHNICAL DRAWINGS Wiring Diagram GH3 300 kg 350 kg 375 kg drive WiFi ...
Страница 145: ...WiFi Guldmann GB US 09 2016 550376_3 147 TECHNICAL DRAWINGS Wiring Diagram GH3 Twin 250 kg 500 kg CLM WiFi ...
Страница 146: ...WiFi Guldmann GB US 09 2016 550376_3 148 TECHNICAL DRAWINGS Wiring Diagram GH3 Twin 250 kg 500 kg load cell WiFi ...
Страница 150: ...Leader PWB Follower PWB Aux Com Guldmann GB US 09 2016 550376_3 152 TECHNICAL DRAWINGS Jumper settings GH3 GH3 ...
Страница 151: ... Guldmann GB US 09 2016 550376_3 153 TECHNICAL DRAWINGS Jumper settings GH3 GH3 Twin ...
Страница 152: ...Leader PWB Aux Guldmann GB US 09 2016 550376_3 154 TECHNICAL DRAWINGS Jumper settings GH1 GH1 Q and GH1 F ...
Страница 154: ... Guldmann GB US 09 2016 550376_3 156 TECHNICAL DRAWINGS Electrical diagram Combi lock 3 single electrical ...
Страница 158: ... Guldmann GB US 09 2016 550376_3 160 TECHNICAL DRAWINGS ...
Страница 159: ... Guldmann GB US 09 2016 550376_3 161 CONTROL Checklist 162 Hoist and rail systems 162 Slings 164 ...
Страница 164: ... Guldmann GB US 09 2016 550376_3 166 NOTES ...
Страница 165: ... Guldmann GB US 09 2016 550376_3 167 NOTES ...
Страница 166: ... Guldmann GB US 09 2016 550376_3 168 NOTES ...
Страница 167: ... Guldmann GB US 09 2016 550376_3 169 NOTES ...
Страница 168: ... Guldmann GB US 09 2016 550376_3 170 NOTES ...