13
9.1 Processing limit
The Injecting Needle is a one-time use product. But follow the steps to
clean, disinfect and sterilize before use.
9.2 Initial processing
9.2.1 Processing principles
It is only possible to carry out effective sterilization after the completion
of effective cleaning and disinfection. Please ensure that, as part of your
responsibility for the sterility of product before use, only sufficiently
validated equipment and product-specific procedures are used for
cleaning/disinfection and sterilization, Please also observe the applicable
legal requirements in your country as well as the hygiene
regulations of the hospital or clinic, especially with regard to the additional
requirements for the inactivation of prions.
9.2.2 Post-operative treatment
The post-operative treatment must be carried out immediately, no later
than 30 minutes after the completion of the operation. The steps are as
follows:
1.Remove the injection needle , please refer to section 5.2.
2.Remove the remaining Gutta Percha materials in the heating cavity . For
specific operations, please refer to point 11 in section 1.2.
Warnings:
The injection needle after surgery cannot be used again.
9.2.3 Preparation before cleaning
Tools: Wrench ,tray, clean and dry soft cloth. The steps are as follows:
1.Installing the injection needle Refer to section 5.2.
2.Squeeze out the Gutta Percha materials in the heating cavity and ensure
that the Gutta Percha materials injected from the Gutta Percha Injecting
Needle exceed 30mm.
3.Remove the Gutta Percha Injecting Needle from the handle with the
wrench provided by guilin woodpecker medical instrument co., LTD. Then
Put them into a clean tray.
4.Wipe and Clean the surface of Gutta Percha Injecting Needle with
a clean cloth dipped in cleaning agent until no dirt can be seen on the
surface. Then dry it with a soft cloth and put them into a clean tray.
Cleaning agent can be pure water.
Notes:
The pure water temperature should not exceed 45 °C, otherwise the protein
will solidify and it is difficult to remove.
9.3 Cleaning
The cleaning should be performed no later than 24 hours after the
operation.The cleaning adopt automated cleaning.
The cleaning procedure are as follows.