10
The contents of the mode that can be adjust includes:
①
Set preset temperature 100-200°C adjustable, 100°C, 120°C, 150°C,
180°C, 200°C);
②
The speed of injecting (fast, medium and slow three gears adjustable);
③
Change the gutta-percha (YES/NO) option, after selecting “YES” ,short
press the injecting button to start the gutta-changing procedure, short press
the “ON/OFF” button to exit the gutta-changing procedure .The specific
steps as folows:
1) Start heating, the display shows “Heating . . . Please wait” , wait for the
temperature to rise to the preset temperature;
2) Start the gutta-percha injection, empty the remaining gutta-percha, the
speed indicator shows the fast gear and flashes until the gutta is emptied.
3) Reset in the fast gear, during the process, the speed indicator on the
display points to the return direction and flashes untile the resetting is
complete;
4) The display shows the prompt “Insert Gutta-Percha” to insert a new
gutta-percha;
5) Press any button to return to the standby interface.
6.3.3 Function setting
In the shutdown state , long press the mode button and the “ON/OFF”
button to turn on the function setting interface, short press the “ON/OFF”
button to adjust the parameters ,short press the mode button to switch
content, short press the injecting button to save the set parameters and exit
the setting interface to enter normal start-up and standby interface. The
functions that can be set are:
①
Set the automatic shutdown time (5-10 minutes adjustable), the default
is 5 minutes;
②
Set the buzzed volume ,3 levels are available,the default is meduim
volume.
6.4 Heating and Canal obturation
After selecting the appropriate mode , short press the injecting button to
start heating, if you need to stop heating,short press the “ON/OFF” button.
The contents displayed on the heating interface in the heating state are:
①
The serial number of mode;
②
Preset temperature;
③
Injecting speed;
④
Battery power.The heating indicator flashes when heating starts, and
when the temperature reachs the preset vaule, the heating indicator is
always on.If you need to stop heating,short press the “ON/OFF” button,
and the heating indicator goes out.
Install the Thermal Protector Cap at the connecting part of Gutta Percha
Injecting Needle and Hot Melting and Filling Instrument, and wipe the
filling material from the needle with gauze and alcohol.
Note: The needle starts filling from the bottom of the root canal to reduce