20
Phenomena of troubles
Possible cause
Solutions
The length measurement
indicator only full display near
narrow part of the apical.
Whether the root canal is
occlusion?
The display will be normal
after penetrating the narrow
part of apical.
If the root canal is too dry?
Wet the endo with Hydrogen
peroxide or NaCl.
If the endo file is too small for
a large root canal?
Replace the current endo file
with a larger one.
* If all above measures do not work, please contact us.
6 Cleaning and sterilization
6.1 The main unit and the measuring wire can be disinfected with alcohol cotton or
disinfected cloth.
6.2 Do not use potent chemical, cleaning agent, etc. to clean the equipment. It will
cause damage to the main unit and the measuring wire.
6.3 The file clip, lip hook and measuring file should be autoclaved under high
pressure (0.22MPa) and high temperature (135
℃
) before each operation.
7 Storage, maintenance and transportation