6
clean tray.
Notes
a) The water used here must be pure water, distilled water or
deionized water.
7.2 Preparation before cleaning
Steps
Tools: tray, soft brush, clean and dry soft cloth Remove
optical fiber from main unit and put it into the clean tray.
Use a clean soft brush to carefully brush the optical fiber until
the dirt on surface is not visible. Then use soft cloth to dry the
optical fiber and put them into a clean tray. The cleaning agent
can be pure water, distilled water or deionized water.
7.3 Cleaning
The cleaning should be performed no later than 24 hours after
the operation.
The cleaning can be divided into automated cleaning and
manual cleaning. Automated cleaning is preferred if conditions
permit.
7.3.1 Automated cleaning
•The cleaner is proved to be valid by CE certificationin
accordance with ENISO 15883.
▪There should be a flushing connector connected to the inner
cavity of the product.
▪The cleaning procedure is suitable for the product, and the
irrigating period is sufficient.
It is recommended to use a washer-disinfector in accordance
with EN ISO15883. For the specific procedure, please refer
to the automated disinfection section in the next section
“Disinfection”.
Notes
a) The cleaning agent does not have to be pure water. It can