Section 4
TM9211B-A-00
6 November 2013
4-12
4.7.5
Calibration of 0.1 A and below (1 ) Ranges
(a)
Remove the Range Extender cable and connect leads from the 10A to the Bridge directly.
(b)
Move the two Range Plugs to 0.1 A. Allow the UUT to stablize 10 minutes prior to starting the
measurements
(c)
Output the first current (starting with lowest current) for the 0.1 A and after proper bridge
measurement cycle (refer to user manuals for bridge), record the measured value under column
(d)
Calculate Actual Deviation and compare to Maximum Deviation. Unit should be within
maximum deviation values. If not, unit will require repair.
(e)
Repeat for all currents listed for each test point. Always start with lowest current (minimum heat
dissipation) and work to highest current.
(f)
Repeat Steps b through d for remaining values.
Note: Measurement Tips.
Consider the following when setting up the measurement
Verify that the max voltage or current applied in the measurement will not exceed the specs for the UUT or
the STD.
Verify the reversal rate is appropriated for the measurement and the uncertainty desired.
Ensure that you know whether the measurement you are reading on the bridge is either a ratio or actual
ohms value.
If using a PC, verify that the number of samples and logging delay are appropriate.
If using a PC set the environmental parameters in BridgeWorks .
Verify guard and ground connections (see 6622A Manual).
If using a scanner, ensure that the proper channels for Rx and Rs are selected.