Section 4
TM9211B-A-00
6 November 2013
4-5
4.6
Nominal Value Calibration Procedure
4.6.1
Record the UUT Information on Table 4.
(b)
Allow the 9211B to stabilize at Calibration environment for a 24 hour period. Ensure that the
room is properly ventilated to dissipate heat up to 30 W of power for the calibration process.
(c)
Setup DCC Bridge and Range Extender and other standards used for appropriate measurement
(refer to Standards Manuals used).
4.6.2
Calibration of 0.01 mA (10 k ) Ranges to 1 A (0.1 Ranges)
(a)
Connect the the 6622A Series Bridge directly (C1 and C2) to the Common and Potential terminals
on the 9211B.
Note: The cables will remain connected in this configuration for the remainder of the calibration.
(b)
Move the two Range Plugs to 0.01 mA. Allow the UUT to stablize 10 minutes prior to starting the
measurements
(c)
Output the first current (starting with lowest current) for the 0.01 mA range and after proper
bridge measurement cycle (refer to user manuals for bridge, record the UUT value under
MEASURED VALUE on the worksheet.
(d)
Unit value should be between the minimum and maximum values. If not, unit will require repair.
(e)
Repeat for all currents listed for each test point. Always start with lowest current (minimum heat
dissipation) and work to highest current.
(f)
Repeat Steps b through d for remaining values.
4.6.3
Calibration of 1 A (100 m ) Range
(a)
Disconnect the current (C1 and C2) lead sets and connect a cable
capable of handling 1 A
ut Common
Terminals. Connect the other end to the 6623A Range Extenders 2 A
Input Common cable connection to the high and low terminals.
Potential Cables should remain connected as before.
(b)
Move the two Range Plugs to 1 A. Allow the UUT to stablize 10 minutes
prior to starting the measurements
(c)
Output the first current (starting with lowest current) for the 1 A and after proper bridge
measurement cycle (refer to user manuals for bridge), ), record the UUT value under MEASURED
VALUE on the worksheet.
(d)
Unit value should be between the minimum and maximum values. If not, unit will require repair.