10
Qualified Printers
•
Implanted device fault
•
Lead fault
•
Battery fault
If you see any of these messages, contact the physician who follows this patient’s
implanted device.
NOTES:
•
The availability of each report feature is dependent upon the device interrogated.
Refer to the heart failure device’s system guide for information about available
features or call Guidant Technical Services. See “Contacting Guidant” on page 5.
•
If the HF PARTNER device is unable to interpret the information from the
implanted device, the record will show ???. Interrogate the implanted device
using a Guidant programmer to receive more information. See “Contacting
Guidant” on page 5.
•
The HF PARTNER device can store up to 10 records. When the HF PARTNER
device memory is full, the oldest record will be deleted to allow the newest record
to be stored. The delete light
will stay on for three seconds.
Qualified Printers
The HF PARTNER system will be provided with a tested and qualified printer. To
verify compatibility of other printers, please call Guidant Technical Services. See
“Contacting Guidant” on page 5.
How to Delete Diagnostic Records
Use the following steps to delete all of the diagnostic records stored in the
HF PARTNER device’s memory: