Operating display:
a) with no KMG connection: LED at green
b) charging process:
LED at red
c) charging process completed, holding charge: LED at green for 19 secs and at red for 1
sec.
The average power consumption in the case of an unlit display is approximately 20 mA,
and in the case of a lit display approximately 40 mA. This gives an operating life of approx-
imately
80 hours in normal display mode and approx . 40 hours in the case of a lit display.
6.2 Storage Space
Whenever the KMG switches to the “ready to measure”
mode and the capacity of the memory is exhausted, the
following message is displayed:
After actuating F3 branching off to the options menu
occurs
Tip
The device has a measured-value memory capable of storing a total of 500 graphs.
This means:
a)
166 normal measurements @ 3 individual measurements
b)
All 500 ‘normal measurements’ @ 1 single measurement
c)
Other combinations of normal/individual measurements
6.3 Memory Space Empty
If, on the menu in 5.2.3,
Check on Measured Values
Held in Memory
, the memory flicks through and
no data is found in the memory bank, the following
message appears:
6.4 Device Still Loaded
In case the KMG is still loaded with a force higher
than 50N if the measurement is started manually the
display will indicate the following message:
By pressing the key F3 the measurement can still be started ignoring the current load.
However in this case the measurement accuracy will not be granted!
If this warning is indicated without a load on the measurement area, the force gauge
should be sent for maintenance to the GTE service department.
Memory empty
Exit
F3
Numbers
exhausted
Options
F1
Device still loaded
Exit
F3