GSI AudioStar Pro™ Clinical Audiometer
84
D-0100778 Rev C
Load:
Allows the operator to select a specific protocol from a list of saved
configurations. This may include back-up configurations or site-specific
configurations.
Save:
Saves selections and settings from the configuration application to a
specific location. This saved configuration may be downloaded at a different time
or to multiple AudioStar Pro audiometers.
Instrument
Security Tab
A list of examiner names and examiner passwords may be entered under the
Instrument/Security tabs of the Config App software. Examiner Passwords are
user defined and may contain any combination of lower or upper case letters and
numbers.
Facility Tab
Facility name, address and logo may be configured from this tab. Date format
and calibration reminders may also be customized. The device regional settings
and the information for 4 speakers, if used, may also be defined in this section.
Please see the AudioStar Pro Config App manual for further information.
Printout Tab
Report preferences are determined by the selections made in this tab. The high
frequency print format, graph orientation, printer protocol, speech printing and
facility logo are customizable items on the printout.
Word Lists Tab
When uploaded from the AudioStar Pro, this window displays the existing word
lists. External CD names may also be added. Word lists may be deleted from the
instrument and Favorites for the word lists are defined in this tab.
Log Tab
In the event of a repeatable error, the
log
window allows the examiner to upload
or email a file from the AudioStar Pro to the computer. This file “retraces your
steps” (button pushes) for the purposes of troubleshooting.
Audiometry
General Tab
Select the frequencies used for PTA calculations, indicate the start-up test mode,
and graph orientation from the general tab. Additionally, the patient response
switch and bone conduction protocol strategies may be determined from this tab.
The routing behavior, reliability label text, auto Hz advance frequency order and
the aux intercom button behavior may also be customized.
Pure tone Tab
Pure Tone Channel 1 and Channel 2 defaults for the start-up stimulus, transducer,
starting intensity and routing defaults are defined in this tab. It is also possible to
assign signal format and dB step size from this tab.