RUN
A scrolling display shows the units selected from the
UNITS/MODE
column, the mode of operation (
Range
,
Level
or
Flow
), the full scale value for the large numeric display,
Totalizer value (in
FLOW
mode), and air temperature at the sensor
location.
LEVEL READING
From RUN use
È
to view the level measured by the instrument.
ECHO CONFIDENCE
Echo Strength is automatically adjusted by the SLT32 in response to
operating conditions in the applications.
From RUN use
È
to get to the Echo Confidence display,
EC
.
TOTALIZER
The Totalizer display will only be enabled when the SLT32 is calibrated in
FLOW
mode. From
RUN
use
È
or
Ç
keys to display
Tot:
value. The Totalizer value is updated every 2 seconds with flow volume
1 litre (0.264 USG). The display will show up to 10 digits and then restart at 0 automatically.
Press
È
or
Ç
keys to return to
RUN
.
The totalizer can be reset to zero by going
Æ
to
SPECIAL FUNCTION
and
È
to
Reset Tot?
.
RELAY STATUS DISPLAY
Press
È
from
Tot:
to
RELAYS 1,2,3
.
Energized relays will be displayed in reverse type
RELAYS
:
SLT32 Level & Flow Monitor
Manual Series B.11
Page 7
Current
Echo Strength
Arrow
Minimum
Marker
EC 9
T: 9999999999
T: 0283168470
TOT:
0
TOT:
0
Units: ft
³, USG, IG, L
Units = m
³
PASSWORD: 00
RUN
Level:
2.000 ft
FLOW MODE ONLY
RELAYS:1 2 3 4 5 6
Tot:
EC 9