QUICK BENCH TEST:
Connect Sensor to the
T.DUCER
terminals as shown below, then apply Power. When properly connected
a soft clicking can be heard from the sensor and figures will show on the large LCD display. Test
operation of the SLT32 by holding the sensor steadily and aiming at a flat, stable target 12 to 28" (305 to
711 mm) away from the end of the sensor. Allow a few seconds for the SLT32 to lock onto the target
before displaying its distance. The SLT32 will now display distance in ft or cm (factory calibration).
NOTE: The SLT32 will not detect targets beyond user entered
MaxRg
.
CONNECTIONS: POWER INPUT:
The standard model requires AC power input between 100-130VAC 50/60Hz (2 amp fuse is
recommended). No adjustments are necessary for any voltage in this range. Optional 230VAC requires
power input between 200-260VAC 50/60Hz. (See OPTIONS section of this manual for connection of
optional 12VDC or 24VDC power input).
IMPORTANT NOTE: To comply with CSA/NRTL standards, AC power input and relay connection
wires must have conduit entry to the instrument enclosure.
SLT32 Level & Flow Monitor
Manual Series B.11
Page 4
T.DUCER
N.O C N.C N.O C N.C N.O C N.C
RELAY 1
RELAY 2
RELAY 3
L N G
AC
RX PU GND TX
RS 232
N.O C N.C N.O C N.C N.O C N.C
RELAY 4
RELAY 5
RELAY 6
4\20
GROUND STUD
:
ATTACH CABLE
SHIELDS (4-20mA, RS232)
THIS END ONLY
IMPORTANT
IMPORTANT:
MUST CONNECT TO A
GOOD GROUND (<1 Ohm)
WITH 12 AWG CONDUCTOR
OPTIONAL
100-130VAC (50/60 Hz)
OPTIONAL:
200-260VAC (50/60 Hz)
*
*
A
L
U
M
IN
U
M
C
H
A
S
S
IS