background image

H75.0.01.6C-02 

 

User’s guide GPH 114 

Page 5 of 8 

Problems may also occur when taking measurements in low-ion media containing solvent. Some of the 
problems occurring when taking measurements in such media can be counteracted by using a double-
chamber electrode 

(Type GE 103) 

with suitable bridging electrolyte (type depending on application).  

4.3  Selection of electrodes 

Special electrodes for various applications: 

1.

Measurements in low ion media 

(rain water, water in aquariums, de-mineralized water)

our type GE 104 (special faceted electrode for > 50

μS/cm) or GE 106 (> 100μS/cm).

2.

Sea water aquariums

Standard combined measuring and reference electrodes 

(GE 100, GE 114).

3.

Swimming pools

Standard pH electrodes 

(GE 100, GE 114).

4.

Soil examinations

Glass electrode with several diaphragms 

(GE 101)

. Use prick lance first!

5.

Cheese, fruit, meat

Insertion electrode 

(GE 101)

.

 

When conducting measurements in cheese, milk and other products

containing proteins, a special cleaning agent has o be used for cleaning of the electrode 

(pepsin solution

- GRL 100)

.

6.

Photographic laboratories

Use double chamber electrode, with jumpering electrolyte (1 molar potassium nitrate solution); potassium
nitrate solution has to be replaced as and when required. Fill watering cap with potassium nitrate solution
for storage of electrode. 

(GE 103)

Normal cleaning

: 0.1 molar HCI solution for at least 5 min. or protein cleaning agent. 

Generally speaking, the service life of electrodes is 8 to 10 months; with proper care and maintenance their 
service life may even be extended to over 2 years. The actual service life is, however, dependant on the 
individual application If you can no longer set the pH X-value this could either be an indication that 
a) the electrode is worn out and needs to be replaced, or that
b) the buffer solution is used up (prepare new solution). Even if treated very carefully during calibration (to
avoid carrying over of buffer solution residuals from one solution to another if electrode is not sufficiently 
cleaned and dried) buffer solutions only keep for a certain period of time (approx. 1 month). 
Buffer capsules can be kept for an unlimited period of time - we, therefore, recommend keeping a certain 
number on stock. 
PH12-buffer-capsules (white) have to be stored in an exsiccator or stored together with drying agent. 
We also recommend keeping a certain amount of 3 mol/l KCl on stock for re-filling. 

4.4  General maintenance an measuring instructions 

This pH electrode has been subjected to a series of tests demanding maximum quality standards in all 
stages of manufacture. 

Attached electrode can be used for 0 to 14 pH respectively at temperatures 

between 0 and 90°C and conductivity > 200 

μS/cm. 

1. Make sure to observe the following points to maintain optimum capacity and accuracy of electrode

as long as possible: 

1.1. Remove storage bottle from pH-electrode and rinse shaft and pH-glass diaphragm with ordinary tap 

water. Then dry shaft with soft tissue. 

1.2. Important! Make sure to always keep pH-glass diaphragm in a slightly moist condition. If electrode is not 

used, the pH glass diaphragm has to be immersed into a 3 mol/l KCl solution for storage. 
Drying out of the pH-glass diaphragm will affect both its capacity and sensitivity. In order to wet it 
throughout, put glass diaphragm in a 3 mol/l KCl solution for 24 hours. 

1.3. Please avoid touching of the glass diaphragm as even the slightest damage to, or abrasion of its surface 

may negatively affect the capacity of the electrode. 

1.4. Make it a rule to always visually check pH-electrodes for any air bubbles that might be enclosed in the 

pH-glass diaphragm and the external reference electrode cell. In case of air bubbles being apparent, 
these can be removed by shaking the electrode (downwards as for a clinical mercury thermometer) 

1.5. During measurements the lateral diaphragm should also come into contact with the material to be 

measured. Minimum immersion depth for GE 114 is 20 mm, max. 50 mm 

1.6. Electrode cable and plug should always be clean and dry as otherwise the electrical insulation may be 

damaged, this resulting in incorrect measurements as well as other faults. 

Содержание GPH 114

Страница 1: ...ese instructions before use Please consider the safety instructions Please keep for future reference PCE Instruments France EURL 76 Rue de la Plaine des Bouchers 67100 Strasbourg France Tel 33 0 972 3...

Страница 2: ...ving from operation must be done by qualified specially trained staff that have carefully read and understood this manual before starting any work The manufacturer will assume no liability or warranty...

Страница 3: ...ay cause in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 2 If there is a risk whatsoever involved in r...

Страница 4: ...ed The picture on the right shows an electrode without temperature sensor There are several design types for the diaphragm but generally said it is the connection between electrolyte and the measured...

Страница 5: ...ctrode is not sufficiently cleaned and dried buffer solutions only keep for a certain period of time approx 1 month Buffer capsules can be kept for an unlimited period of time we therefore recommend k...

Страница 6: ...ne temperature of buffer solution and set accordingly using the central turning knob controller for temperature compensation 0 to 90 C 1 graduation mark approx 10 C Place GPH 114 meter on a solid surf...

Страница 7: ...ransport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 7 2 Disposal instructions Batteries mus...

Страница 8: ...and the dilution influence dpH are values to measure this capability The buffer capacity is the amount of a strong acid or base that has to be added to 1 liter of the buffer solution in order to chang...

Отзывы: