GREISINGER GPH 114 Скачать руководство пользователя страница 6

H75.0.01.6C-02 

 

User’s guide GPH 114 

Page 6 of 8 

2. Care and maintenance:

2.1. Crystallization of the 3 mol/l KCl is not avoidable! 

Crystallized KCl at protection cap or closing collar can easily be removed by fingernail or cloth and is no 
defect or reason for reclamation. 

2.3. Dirty electrodes have to be cleaned. You will find suitable cleaning agents for the pH-glass diaphragm in 

the following table: 

Contamination  

Cleaning agent 

Various deposits  

Light cleaning agent 

Inorganic coatings  

Commercial liquids for cleaning of glass 

Metal compounds  

1 mol/l HCI solution 

Oil, grease  

Special cleaning agents or solvents 

Biological coatings containing proteins 

1 % pepsin enzyme in 0.1 molar HCI solution 

Resin-lignines  

Acetones 

Highly resistant deposits  

Hydrogen superoxide, sodium hypo chloride 

The distinct materials of the electrode have to be considered when choosing the cleaning agent. 
E.g. electrodes with plastics shaft must not be cleaned with solvents. In case of doubt please contact the 

manufacturer for further advice. 

The same has to be considered when using aggressive or other, non water based agents! 

5  Calibration of pH-Meter 

Accessories required: calibrating solution for pH 7 und pH 4 (resp. special accessories pH 10, pH 12) 

How to prepare a calibrating solution: chapter 10 Notes B: preparation of pH buffer solutions  

5.1  Calibration of pH-Meters 

In order to ensure optimum measuring accuracy, the calibration range should be larger than the measuring 
range. We recommend using the following calibration solutions for measurements. 
less than pH 7: pH 4,0 and pH 7,0 
more than pH 7: pH 7,0 and pH 12,0 
Plug in Cinch plug of pH-electrode into corresponding socket and switch on device by means of the lateral 
slide switch (a number with decimal point will be displayed on the LCD). 
Determine temperature of buffer solution and set accordingly using the central turning knob (controller for 
temperature compensation 0 to 90°C): 1 graduation mark = approx.10°C. 
Place GPH 114 meter on a solid surface; if possible try to avoid holding the meter in your hand during 
calibration (please refer 

to “Points to be observed during measuring”). 

5.2  Setting of the first calibration point 

Carefully remove protection cap from electrode. (Attention: cap contains 3 mol/l KCl). 
Rinse electrode with distilled water, dry electrode, place it in buffer solution pH 7.0. 
Wait approx. 20 to 30 seconds (for display to show stable value), then use turning knob at the far right side 
(pH 7) to set value to 7.00. 

5.3  Setting of second calibration point 

In order to check calibration repeat procedures for calibration points 1 and 2 using corresponding turning 
knob for re-adjustments. 
In case you did not prepare pH 4.0 but another buffer solution, e.g. pH 10.0 or pH 12.0, the corresponding 
value (pH 10.0 or pH 12.0) can be set (calibration of second calibration point) using the left-hand turning 
knob (pH X). 
Please note that for each and every calibration procedure the buffer solution value pH 7.0, i.e. 7.00, has to 
be set by means of the right-hand turning knob (ph 7). 

Note: 

We recommend to re-calibrate the measuring device prior to each new measuring series to 
ensure optimum accuracy. It has to be considered that the calibration solution and the 
media to be measured should had similar temperatures! 

After completion of measurements fill electrode protection cap with 3 mol/l KCl solution and slip it back onto 
electrode. (Gentle squeezing of the cap displaces air, thus allowing cap to be slipped on easily) 

Attention: 

pH-electrodes are extremely delicate and sensitive measuring elements. Prior to using the pH 
electrode, please carefully read corresponding maintenance and measuring manual! 

We shall not assume any warranty if customer fails to use proper care and skill 
(e.g. breaking of electrode, drying-out, blocking etc.).

  

Содержание GPH 114

Страница 1: ...ese instructions before use Please consider the safety instructions Please keep for future reference PCE Instruments France EURL 76 Rue de la Plaine des Bouchers 67100 Strasbourg France Tel 33 0 972 3...

Страница 2: ...ving from operation must be done by qualified specially trained staff that have carefully read and understood this manual before starting any work The manufacturer will assume no liability or warranty...

Страница 3: ...ay cause in a failure of the function In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start up 2 If there is a risk whatsoever involved in r...

Страница 4: ...ed The picture on the right shows an electrode without temperature sensor There are several design types for the diaphragm but generally said it is the connection between electrolyte and the measured...

Страница 5: ...ctrode is not sufficiently cleaned and dried buffer solutions only keep for a certain period of time approx 1 month Buffer capsules can be kept for an unlimited period of time we therefore recommend k...

Страница 6: ...ne temperature of buffer solution and set accordingly using the central turning knob controller for temperature compensation 0 to 90 C 1 graduation mark approx 10 C Place GPH 114 meter on a solid surf...

Страница 7: ...ransport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 7 2 Disposal instructions Batteries mus...

Страница 8: ...and the dilution influence dpH are values to measure this capability The buffer capacity is the amount of a strong acid or base that has to be added to 1 liter of the buffer solution in order to chang...

Отзывы: