13  Replace sensor element 

Unplug the sensor, unscrew the white protection housing and take out sensor 
element. Remove rubber disks and put it on the new sensor. Put the new sensor 
element into the housing, screw it down and plug it. Then start the calibration. 

14  Specification 

Measuring range:

 

0.0  ... 100.0 %O2 (oxygen concentration) 

MOD (based on O

2

 conc. and max. O2-partial pressure) 

Accuracy:

  

calibrated device at nominal temp.: ± 0.1 %O2  ± 1 digit 

Sensor linearity: < 2 vol.-%: +/- 0.1%; < 25 vol.-%: +/- 
0.5%; < 100 Vol.-%: +/- 1.0% 

Sensor connection 

0.7m connection cable (fixed to device) with jack connector 

 

Sensor: 

electrochemical partial oxygen pressure sensor, integrated 
in sensor housing (type for replacement: GOEL 370)

 

 Response time: 

90% in < 10 sec., depending on temperature 

 Operating life: 

guaranteed 12 months (assuming appropriate usage) 

 Operating pressure:  0.5 to 2.0 bar absolute. (at one-sided strain: max. 0.25 bar 

over-/under-pressure) 

Display:

 

approx. 13 mm high, 3½-digit LCD 

Pushbuttons:

 

3 keys for ON/OFF, min-/max-value display, calibration 

Nominal temp.:

 

25°C 

Ambient temp.:

 

0 to 45°C (sensor), -20 to 50°C (device) 

Relative humidity:

 

0 to 95 %RH (not condensing) 

Storage temperature:

 

-15 to 60°C (sensor), -20 to 70°C (device) 

Power Supply:

 

9V-battery type 6F22 (in scope of supply) 

  Power Consumption: 

0.14mA (standard zinc carbon battery: >2100 hours!) 

  Battery Change Indicator:  automatically if battery is used up: "BAT" 

Auto-Power-Off:

 

when activated, the device switches automatically off, if it is 
not operated for longer time (selectable 1..120min). 

Housing: 

  Dimensions: 

impact-resistant ABS, transparent panel, front side IP65 
approx. 106 x 67 x 30 mm (L x W x D) without sensor cable 

Weight:

 

approx. 185g  incl. battery and sensor 

EMC

  

The device corresponds to the essential protection ratings 
established in the Regulations of the Council for the 
Approximation of Legislation for the member countries 
regarding electromagnetic compatibility (2004/108/EG). 
Device meets EN 61326-1:2006, Additional fault: <1% 

Содержание GOX 100T

Страница 1: ...k GmbH Standort Greisinger Hans Sachs Stra e 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Oxygen Measuring Device for diving gas mixtures WEEE Reg Nr DE 93889386 as of V...

Страница 2: ...s will be detected at calibration and the sensor element has to be replaced before the next measurement By means of the integrated hold function the MOD Display max operating depth and a specialized s...

Страница 3: ...y to avoid re starting Operator safety may be a risk if there is visible damage to the device or the device is not working as specified the device has been stored under unsuitable conditions for a lon...

Страница 4: ...has to be replaced measuring is valid On Off switch Press shortly change between actual value oxygen hold value oxygen hold value MOD in meters n or feet F Press shortly short display of the sensor ra...

Страница 5: ...out at other pressure conditions than the calibration The actual pressure has to be entered in the configuration for measurement and calibration The sensor s temperature dependency is compensated inte...

Страница 6: ...Press shortly the key the display shows for a short time i e 100 The rating is displayed in 10 percent steps 100 optimal sensor status Lower values are indicating that the sensor life time will soon b...

Страница 7: ...e 1 to 120 min OFF auto power off is deactivated continuous operation Abs air pressure factory setting 1013 mbar 500 1200 Input of absolute pressure of measured gas The absolute pressure is depending...

Страница 8: ...ed strain max 0 25 bar over under pressure Display approx 13 mm high 3 digit LCD Pushbuttons 3 keys for ON OFF min max value display calibration Nominal temp 25 C Ambient temp 0 to 45 C sensor 20 to 5...

Отзывы: