12 Configuration of the device
Follow this instructions to configure the functions of the device:
Switch the device off.
Press
mode
-button and keep it pressed, while turning the instrument on.
Release the mode button after the segment test (
) .
The first parameter
is displayed .
If a parameter should be changed press key up or down ( ), the present
setting of the parameter will be shown and can be changed via up/down keys.
Jump to the next parameter by pressing
.
Parameter Value
Information
button
buttons
Auto Power-Off
(turn-off delay)
factory setting: 20 min
1
…
120
Turn-off delay in minutes. If no key is pressed for the time
adjusted in this parameter, the device is automatically switched
off to safe battery power. (adjustable range 1 to 120 min)
OFF
auto power-off is deactivated (continuous operation)
Abs. air pressure
factory setting: 1013 mbar
500
…
1200
Input of absolute pressure (of measured gas). The absolute
pressure is depending on altitude above sea level (refer to
following table) and on the weather.
Max. oxygen partial pressure for MOD display
factory setting: 1.40 bar
1.00
…
1.60
Input in bar
Unit for MOD display
factory setting:
n
N
meters
F
feet
Pressing again stores the settings, the instrument restarts.
Please note: If there is no key pressed within the menu mode within 2 minutes, the
configuration will be cancelled, the entered settings are lost!
Air pressure depending on the altitude above sea level
Altitude
Abs. air pressure
Altitude
Abs. air pressure
0 m
1013 mbar
800 m
921 mbar
100 m
1001 mbar
1000 m
899 mbar
200 m
989 mbar
1200 m
877 mbar
300 m
978 mbar
1400 m
856 mbar
400 m
966 mbar
1600 m
835 mbar
500 m
954 mbar
1800 m
815 mbar
600 m
943 mbar
2000 m
795 mbar