10  Error And System Messages  

Display 

Description / Reason 

Remedy 

 

No sensor  
 
or sensor defective 

Connect sensor  
 
sensor defective => return sensor to manufac-
turer for repair 

 

Low battery voltage  
device will only continue operation for 
a short time 

replace battery 

 

Low battery voltage 

replace battery 

- If mains operation: wrong voltage 

replace power supply, if fault continues to exist: 
device damaged 

No display

 

or 

mazy characters

 

- Battery voltage too low 

replace battery 

- If mains op.: power supply defective 

or wrong voltage/polarity 

check/replace mains supply 

- System error 

disconnect battery or power supply, wait for a 
short time, re-connect 

- Device defective 

return to manufacturer for repair 

 

Values exceeding measuring range 
 

Are there any values exceeding the measuring 
range specified? 
 ->measuring value too high 

Sensor/cable defective 

-> replace probe 

 

Values below measuring range 

Are there any values below the measuring 
range specified?  
-> measuring value too low  

Sensor/cable defective 

-> replace probe 

 

System fault 

switch on again: if fault continues to exist, de-
vice is damaged  
-> return to manufacturer for repair 

 

No sensor 

connect suitable sensor 

Error in sensor 

Sensor defective => return to manufacturer for repair 

Temperature display correct,  
oxygen display incorrect 

check: plug in sensor housing connected? 
Open PG-glanding and pull up plug as far as 
possible. 

 

Value could not be calculated 

One measuring variable required for calculation 
is missing (no sensor) or incorrect (over-
flow/underflow) 

10.1  Error and System Messages during Oxygen Calibration 

Display 

Description / Reason 

Remedy 

(AL 

ERR. 1 

Wrong temperature 

temperature has to be between 5 and 40°C 

(AL 

ERR.3 

Sensor slope to low 

check calibration environment (p.r.t. 'Calibration 
of the oxygen sensor'). If error are unremedied 
replace sensor element. 

(AL 

ERR.4 

sensor slope to high 

check calibration environment  
(p.r.t. 'Calibration of the oxygen sensor') 

(AL 

ERR.5 

calculated offset to large 

check calibration environment (p.r.t. 'Calibration 
of the oxygen sensor'). If error are unremedied 
replace sensor element. 

Содержание GMH 3691

Страница 1: ...93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device with alarm function for gaseous oxygen and temperature as of version V1 6 Operat...

Страница 2: ...AGES 9 10 1 ERROR AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 9 11 THE SERIAL INTERFACE 10 11 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 10 12 SPECIFICATION 11 13 DISPOSAL INSTRUCTION 11...

Страница 3: ...e operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device 1 Trouble free operation a...

Страница 4: ...xygen and temperature values have been stable for a longer period of time CAL arrow indicates that an automatic oxygen calibration is carried out Alarm arrow indicates an alarm by blinking Warning tri...

Страница 5: ...e probe The mains socket is located at the left side of the instrument 6 Configuration For configuration of the device press key for 2 seconds the main menu of the configuration will be called up Use...

Страница 6: ...ower off delay in minutes Device will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place automati cally deactivated for cyclic l...

Страница 7: ...asuring please note When measuring gases please consider the following Calibration and measuring are depending of the absolute pressure at the sensor Therefore check absolute pressure before calibrati...

Страница 8: ...lease note If You started editing the manual value once the change to air is no more possible 3 Calibration point 1 Pt 1 Expose sensor to the chosen gas and wait until temperatures of gas and sensor h...

Страница 9: ...below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continues to exist de vice is damaged return to manufacturer for...

Страница 10: ...ablished programming languages suitable for Windows XP Windows Vista Windows 7 Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint 11 1 The following interface functions will be supported Chann...

Страница 11: ...prox 2 mA Automatic off function Device will be automatically switched off if no key is pressed no interface communica tion takes place for the time of the power off delay The power off delay can be s...

Страница 12: ...H36 0 12 6C 09 Operating Manual GMH 3691 page 12 of 12...

Страница 13: ...ing chambers for cell cultures Monitoring of fermen ting process of fruits in fermentation plants etc Medicine Monitoring of oxygen concentration in respirators checking of breathing monitoring of oxy...

Отзывы: