50

WK

Handheld instrument

Display / Controller

Logger / EASYBus

Transmitter

Temperature probe

Alarm / Protection

Air oxygen measuring device

  Double display for oxygen and temperature

  Measured units: O

2

-concentration and 

O

2

-partial pressure

 

Alarm detector with integrated horn

  Automatic temperature compensation

  Min./Max. value memory, Hold function

  Serial interface, device can be connected 

to bus system (up to 5 devices can be con-

nected to one PC interface)

  Battery and d.c. operation

  Wide range of application

  Most simple calibration in atmospheric air

GMH

 

3691 

 

Sensor not included - please order separately!

 

Residual oxygen meas. device

for quick and cost-effective measure-

ment of residual oxygen

GMH

 

3691

 

GOG 

 

Application:

Essentially  there,  where  delicate  products  are 

conserved by low-oxygen atmospheres (protec-

tive gas), this instrument is suitable to check the 

residual oxygen content.

 

  packaging industry

 

  food industry

Specification:

 

(summary)

Meas. range:

 0,0 ... 100,0 % O

2

 

(O

2

-concentration)

Accuracy:

 

(whole system - during carefully 

calibration and measuring)

 1-point-calibration:  ±0.2 %O

2

 ±1 digit

                

(for concentrations < 10%) 

 2-point-calibration:  ±0.1 %O

2

  ±1 digit  

                

(for concentrations < 10%)

Oxygen probe: 

 Oxygen-partial pressure probe, 

built in external sensor housing

Response time:

  T

90

 < 10 sec., depending on 

temperature

Operation life:

 

warranty for sensor element 

12 months 

(appropriate appli-

cation and ambient pressure)

Working pressure:

 0.5 to 2.0 bar abs.

Over-/under-pressure:

 max. 0,25 bar 

Working temperature:

  0 to 50°C (sensor), 

                     -20 to 50°C (device)

Relative humidity:

  0 to +95%RH 

(non-condensing)

Storage temperature:

  -15 to 60°C (sensor), 

                     -20 to 70°C (device)

Power supply:

  9V battery type IEC 6F22

Dimensions case:

 approx. 394 x 294 x 106 mm

Weight:

  approx. 1400g  (cpl. set)

for additional technical data refer to 

GMH3691 and accessory sensors p. 51

Scope of supply:

 

Instrument GMH3691, hand pump with air tube, 

GOG oxygen sensor with penetration needle, case 

GKK3500,  spare  needle  ø0,9mm,  rubber  foam 

sticker (40 pieces), operating manual.

Spare elements, accessories:

GOG-SET

 

Set without instrument 

  

Scope of supply: GOG oxygen sensor with 

penetration needle, hand pump with air tube, case 

GKK3500, spare needle and 40 rubber foam sticker

GOEL 370

 

spare sensor element 

  

 

GOG-N

 

needle, 

Ø

 

0.9

 

mm (5 pieces)

 

 

GOG-A

 

rubber foam sticker (40 pieces)

 

 

ST-R1

 

device protection bag

 

 

 with cut-out for probe connection

 for add. accessories  p.r.t. page 56 - 58

Calibration:

  1-point calibration: extremely 

simple quick calibration in atmospheric air. (press 

button to compensate unit to 20.9%).

2-point calibration: first point at atmospheric air 
(20.9%), second point freely selectable

Application:

 

Wide range of application for your home, job and 

hobby! For example:

- Bio chemistry:

 Oxygen monitoring in breeding 

chambers for cell cultures. Monitoring of fermen-

ting process of fruits in fermentation plants etc.

- Medicine:

 Monitoring of oxygen concentration 

in respirators; checking of breathing, monitoring 

of oxygen concentration in incubators, oxygen 

tents etc.

- Food technology:

 Monitoring of residual oxygen 

in packages (e.g. coffee, tea, etc.). Monitoring of 

oxygen content during production processes.

- Air conditioning and ventilation technology:

 

Oxygen measurements, air quality monitoring, 

measuring of oxygen concentration in enclosed 

air conditioning systems, etc.

- Sport:

 Checking of oxygen content in com-

pressed air breathing apparatuses (diving, etc.), 

oxygen monitoring for gliding.

The device can only be used to check during 

these applications. -> no substitute for ap-

proved monitoring device!

Specification:

Measuring ranges:

       

 

Oxygen concentration:

    0,0 ... 100,0 %

 

O

2

   

                         (gaseous) 

 

Partial oxygen pressure:

   0 ... 1100 hPa O

2

 

Temperature:

    -5,0 ... 50,0 °C

Accuracy:

 

(device)

 

(at nominal temperature = 25°C)

 

Oxygen concentration:

    ±0.1% ±1digit

 

Partial oxygen pressure:

  ±1 hPa ±1digit

 

Temperature:

    ±0.1°C ±1digit

Oxygen electrode:

   for suitable sensores 

                   p.r.t. page 51

Sensor connection:

 6-pin screened Mini-DIN-

socket.

Display: 

two 4 digit LCDs 

(12.4mm or 7mm high)

as well as additional arrows.

Pushbuttons:

 6 membrane keys for ON/OFF-

switch, selection of meas. range, min- and max- 

value memory, hold-function, calibration etc.

Working temperature:

 0 to +50°C 

Relative humidity:

  0 to +95%RH 

(non-condensing)

Storage temperature: 

-20 to +70°C

Interface:

  serial interface,

direct connection to RS232 or USB interface of 

a PC via electrically isolated interface converter 

GRS3100 or GRS3105 resp. USB3100 (p.r.t. 

accessories). 

Power supply:

 9V-battery, type IEC

 

6F22  

(in-

cluded),

 as well as additional d.c. connector for 

external 10.5-12V direct voltage supply. (suitable 

power supply: GNG10/3000)

Power-Off-function:

 1...120min (can also be 

deaktivated).

Power consumption: 

approx. 1.5 mA

Low battery warning:

   and ' bAt '

Dimensions:

 142 x 71 x 26 mm (H x W x D)

Impact-resistant ABS plastic housing, membrane 

keyboard, transparent panel. Front side IP65, 

integrated pop-up clip.

Weight:

 approx. 160 g  (cpl. with battery)

Functions:

Min-/Max-value memory:

 max. and min. values 

will be memorized.

Hold function:

 by pressing a button the current 

meas. value will be memorized.

Alarm:

 integrated limit detector for min. or max. 

alarm. 

Temperature compensation:

 automatic via tem-

perature sensor, integrated in probe housing. 

Air pressure compensation:

 The O

2

 concen-

tration will be compensated according to the abs. 

  atmospheric pressure set (500...2000hPa).

Accessories:

Suitable sensores

 

p.r.t. page 51

GKK

 

3000 

case 

(275 x 229 x 83 mm)

  

 

 with punched lining suitable for GMH3xxx

USB

 

3100 N

 

interface converter,

 

 

 electrical isolated

GRS

 

3105

 

interface converter 

 

 

 with 5 connection points, electr. isolated, for the 

 connection of 5 GMH3xxx to one PC (RS232).

ST-R1

 

device protection bag

 

 

 with cut-out for probe connection
 

 for add. accessories  p.r.t. pages 56 - 58

Содержание GMH 3691

Страница 1: ...93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device with alarm function for gaseous oxygen and temperature as of version V1 6 Operat...

Страница 2: ...AGES 9 10 1 ERROR AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 9 11 THE SERIAL INTERFACE 10 11 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 10 12 SPECIFICATION 11 13 DISPOSAL INSTRUCTION 11...

Страница 3: ...e operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device 1 Trouble free operation a...

Страница 4: ...xygen and temperature values have been stable for a longer period of time CAL arrow indicates that an automatic oxygen calibration is carried out Alarm arrow indicates an alarm by blinking Warning tri...

Страница 5: ...e probe The mains socket is located at the left side of the instrument 6 Configuration For configuration of the device press key for 2 seconds the main menu of the configuration will be called up Use...

Страница 6: ...ower off delay in minutes Device will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place automati cally deactivated for cyclic l...

Страница 7: ...asuring please note When measuring gases please consider the following Calibration and measuring are depending of the absolute pressure at the sensor Therefore check absolute pressure before calibrati...

Страница 8: ...lease note If You started editing the manual value once the change to air is no more possible 3 Calibration point 1 Pt 1 Expose sensor to the chosen gas and wait until temperatures of gas and sensor h...

Страница 9: ...below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continues to exist de vice is damaged return to manufacturer for...

Страница 10: ...ablished programming languages suitable for Windows XP Windows Vista Windows 7 Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint 11 1 The following interface functions will be supported Chann...

Страница 11: ...prox 2 mA Automatic off function Device will be automatically switched off if no key is pressed no interface communica tion takes place for the time of the power off delay The power off delay can be s...

Страница 12: ...H36 0 12 6C 09 Operating Manual GMH 3691 page 12 of 12...

Страница 13: ...ing chambers for cell cultures Monitoring of fermen ting process of fruits in fermentation plants etc Medicine Monitoring of oxygen concentration in respirators checking of breathing monitoring of oxy...

Отзывы: