H36.0.12.6C-09 

Operating Manual GMH 3691 

page 3 of 12

 

 

Operating and Maintenance Advice 

a)  When to replace battery: 

If 

 and ´bAt´ are shown in the lower display the battery has been used up and needs to be replaced. The de-

vice will, however, operate correctly for a certain time. 
If ´bAt´ is shown in the upper display the voltage is too low to operate the device; the battery has been completely 
used up. 

Please note:  The battery has to be taken out, when storing device above 50°C. 

We recommend to take out battery if device is not used for a longer period of time. 

b)  Treat device and sensor carefully. Use only in accordance with above specification. (do not throw, hit against etc.). 

Protect plug and socket from soiling. 

c)  Make sure to use sensors that are suitable for the device. Unsuitable measuring probes may lead to the destruction 

of the measuring device and the measuring probes. 

d)  When connecting the electrode the connector may not lock to the jack correctly. In such a case hold the connector 

not at the case but at the buckling protection of the cable during the plug in. 
Don't connect electrode canted! If plug is entered correctly, it will slide in smoothly. 
To disconnect sensor do not pull at the cable but at the plug 

e)  Mains operation: 

When using a power supply device please note that operating voltage has to be 10.5 to 12 V DC. 
Do not apply overvoltage!! Cheap 12V-power supply devices often have excessive no-load voltage. We, there-
fore, recommend using regulated voltage power supply devices. Trouble-free operation is guaranteed by our 
power supply devices. Trouble-free operation is guaranteed by our power supply, GNG10/3000.  
Prior to connecting the plug power supply device with the mains supply make sure that the operating voltage 
stated at the power supply device is identical to the mains voltage. 

Safety Requirements 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and 
special safety advises given in this manual will be adhered to when using the device. 

1.  Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any 

other climatic conditions than those stated under "Specification". 

2.  If the device is transported from a cold to a warm environment condensation may cause in a failure of the func-

tion. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a 
new start-up. 

3.  If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed most care-

fully. Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible 
voltages impairing or destroying the device or another device connected. 

 

Warning: 

If device is operated with a defective mains power supply (short circuit from mains voltage to output 

voltage) this may result in hazardous voltages at the device (e.g. sensor socket, serial interface). 

4.  If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be 

marked accordingly to avoid re-starting. 

 

Operator safety may be a risk if: 

 

 - there is visible damage to the device 

 

 - the device is not working as specified 

 

 - the device has been stored under unsuitable conditions for a longer time. 

 

In case of doubt, please return device to manufacturer for repair or maintenance. 

5. 

Warning: 

Do not use these products as safety or emergency stop devices or in any other application where failure 

of the product could result in personal injury or material damage. 
Failure to comply with these instructions could result in death or serious injury and material damage. 

6.  This device only serves as supervision by the monitoring of essential or other for the customer important 

systems. 
It must not be used instead of compulsory approval monitoring devices an it is not planed for that purpose.  
If this device is used for the monitoring of such systems on its own, the manufacturer will not assume li-
ability for damages whatsoever. 

 

 

Содержание GMH 3691

Страница 1: ...93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Hand held measuring device with alarm function for gaseous oxygen and temperature as of version V1 6 Operat...

Страница 2: ...AGES 9 10 1 ERROR AND SYSTEM MESSAGES DURING OXYGEN CALIBRATION 9 11 THE SERIAL INTERFACE 10 11 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 10 12 SPECIFICATION 11 13 DISPOSAL INSTRUCTION 11...

Страница 3: ...e operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device 1 Trouble free operation a...

Страница 4: ...xygen and temperature values have been stable for a longer period of time CAL arrow indicates that an automatic oxygen calibration is carried out Alarm arrow indicates an alarm by blinking Warning tri...

Страница 5: ...e probe The mains socket is located at the left side of the instrument 6 Configuration For configuration of the device press key for 2 seconds the main menu of the configuration will be called up Use...

Страница 6: ...ower off delay in minutes Device will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place automati cally deactivated for cyclic l...

Страница 7: ...asuring please note When measuring gases please consider the following Calibration and measuring are depending of the absolute pressure at the sensor Therefore check absolute pressure before calibrati...

Страница 8: ...lease note If You started editing the manual value once the change to air is no more possible 3 Calibration point 1 Pt 1 Expose sensor to the chosen gas and wait until temperatures of gas and sensor h...

Страница 9: ...below the measuring range specified measuring value too low Sensor cable defective replace probe System fault switch on again if fault continues to exist de vice is damaged return to manufacturer for...

Страница 10: ...ablished programming languages suitable for Windows XP Windows Vista Windows 7 Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint 11 1 The following interface functions will be supported Chann...

Страница 11: ...prox 2 mA Automatic off function Device will be automatically switched off if no key is pressed no interface communica tion takes place for the time of the power off delay The power off delay can be s...

Страница 12: ...H36 0 12 6C 09 Operating Manual GMH 3691 page 12 of 12...

Страница 13: ...ing chambers for cell cultures Monitoring of fermen ting process of fruits in fermentation plants etc Medicine Monitoring of oxygen concentration in respirators checking of breathing monitoring of oxy...

Отзывы: