10  Indication of Electrode State 

(for pH-measurements only)

 

The electrode state calculated on basis of the last successful calibration will be displayed 3 seconds by pressing the 

 -key. The state will also be displayed automatically after calibration. 

 

100%:  

optimum electrode state 

30...90%:   satisfactory electrode state 
<30%:  

electrode considerably aged or soiled. Please replace electrode if 
there is no improvement after it has been cleaned and calibrated 
acc. to paragraph ’pH electrode’ 

For the percent evaluation both asymmetry and slope will be taken into account, the lower result will then be used to 
calculate the electrode state. 

Use  

 -key to display electrode characteristics: 

 

Electrode asymmetry at 25°C [mV] 
 

max. permissible range: ±60mV, optimum: 0mV  
Soiling of the electrode has a negative effect on the electrode asymmetry.

 

After pressing the 

 -key once again shortly: 

 

Slope of electrode at 25°C [mV/pH] 

permissible range: -62...-45mV/pH, optimum: -59,2mV/pH 

In case of 2-point calibration the slope will be stated for the entire measuring 
range. For 3-point calibration the slope for the acid measuring range will be 
displayed (SL.1) first. By pressing the 

 -key once again the slope for the ba-

sic range (SL.2) will be displayed 

11  How to Perform a rH-Measurement

 

 

The rH-value of a liquid will be calculated from the measurements of the pH-value, the redox value, and the tem-
perature of a liquid. To determine the rH-value of your solution, proceed as follows: 

Please note: 

Make sure that pH- and redox electrodes are in a perfect condition during measuring and that they are 

cleaned and dried carefully before placing them in the solution.  
First put pH- and redox electrode and the temperature sensor in the solution, stirring it carefully.  

 

1. How to measure pH-value: 

Connect the pH-electrode and the temperature sensor to the GMH 3530. Then 
set GMH3530 to pH-measuring mode and calibrate electrode, if necessary, 
(p.r.t. configuration and calibration during measuring mode pH). 
Then take measurements of the pH-value of the solution and press the 

 -key 

to memorize measurement. Do not switch off the GMH3530 before the rH-
measurement has been completed as otherwise the pH-value could be deleted 
and will have to be entered manually. 
 

 

2. How to establish rH-value: 

Put redox electrode and temperature sensor in the solution, stirring it carefully.  
Connect redox-electrode and configurate the GMH3530 to rH-measuring. The 
main display shows the rH-value of the solution, the secondary display switches 
over between the pH-value measured before and the temperature.  

Please note: 

The measuring values for pH and temperature (if no temperature probe is connected) can also be 
entered manually. Press 

 for a short time and use 

 and 

 to enter temperature value. After 

pressing 

 shortly the pH-value can be changed. (also refer to manual temperature settings). 

 

 

pH

mV

mV

H

rH

ATC

ok

CAL

AL Logg

pH

mbar

pH

mV

mV

H

rH

ATC

ok

CAL

AL Logg

pH

mbar

pH

mV

mV

H

rH

ATC

ok

CAL

AL Logg

pH

mbar

pH

mV

mV

H

rH

ATC

ok

CAL

AL Logg

pH

mbar

pH

mV

mV

H

rH

ATC

ok

CAL

AL Logg

pH

mbar

mbar

mbar

Содержание GMH 3530

Страница 1: ...09 GREISINGER electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for pH Redox temperature as of version V2 6 Operating Manu...

Страница 2: ...AND SYSTEM MESSAGES 10 13 THE PH ELECTRODE 11 13 1 VARIOUS APPLICATIONS REQUIRE SPECIAL ELECTRODES 11 14 THE REDOX ELECTRODE 12 15 THE SERIAL INTERFACE 13 15 1 THE FOLLOWING INTERFACE FUNCTIONS WILL B...

Страница 3: ...s connected or the connection cable is damaged the display will nevertheless show mV pH or rH values Please note that these values can never be correct measuring results 4 Safety Requirements This dev...

Страница 4: ...ed press for 2 sec the min or max value will be deleted Configuration to enter values or change settings CAL for pH mode only press shortly display state of electrode condition and calibration data pr...

Страница 5: ...for pH measurements only 2 Pt the pH electrode will be calibrated at 2 points one calibration point in neutral range and one calibration point in acid or basic range 3 Pt the pH electrode will be cal...

Страница 6: ...measuring mode 8 Manual Setting of pH Value for measuring function rH Running the GMH 3530 in the rH measuring mode requires both temperature and pH value Press shortly again to switch over to the pH...

Страница 7: ...mum calibration of the device 9 3 How to carry out calibration Please note the calibration can only carried out at a temperature range of 0 60 C If you have not yet done so set device to measuring mod...

Страница 8: ...of temperature for solution 2 Use or keys to enter the buffer solution temperature Use to take over value and to display electrode state 5 Rinse electrode in distilled or non ionised water dry electro...

Страница 9: ...ermine the rH value of your solution proceed as follows Please note Make sure that pH and redox electrodes are in a perfect condition during measuring and that they are cleaned and dried carefully bef...

Страница 10: ...eck are there any values below the measur ing range specified meas device not suitable Electrode sensor cable defective replace electrode probe System fault switch on again if fault continues to exist...

Страница 11: ...embrane or the diaphragm will influence the measurements and have to be removed at regular intervals This can be done by means of automatic cleaning equipment 13 1 Various applications require special...

Страница 12: ...mometer Check reference electrolyte level at regular intervals if necessary top up electrolyte level with 3 mol lKCI solution through filling hole closed by means of a silicon ring using a syringe or...

Страница 13: ...ting mode pH mV rH Ther Channel 1 2 1 2 1 2 1 x x x x x x x 0 Read nominal value x x x 1 Set nominal value x x x x x x x 3 Read system status x x x x x x x 6 Read min value x x x x x x x 7 Read max va...

Страница 14: ...elay can be set to values between 1and 120 min it can be completely deactivated Housing dimensions 142 x 71 x 26 mm L x W x D impact resistant ABS plastic housing membrane key board transparent panel...

Отзывы: