H33.0.02.6C-09
Operating Manual GMH 3530
page 14 of 14
16 Specification:
Display ranges:
Temperature:
-100.0 ... +250.0°C or -148.0 ... +482.0°F
pH:
0.00 ... 14.00 pH
Redox (ORP):
-1999 ... +2000 mV;
referring to hydrogen system: -1792 ... +2207 mV
H
(bei 25°C) acc. to DIN 38404
rH:
rH 0.0 ... 70.0
Resolutions:
0.1°C or 0.1°F / 0.01 pH / 1 mV / 0.1 rH
Accuracy:
(at nominal temperature, device ±1Digit)
Temperature:
±0.2°C (-20..80°C), otherwise ±0.4°C
pH:
±0.01 pH (for electrode temperature 10..50°C)
Redox:
±0.1% FS (mV or mV
H
)
rH:
±0.1 rH
Sensor connections:
pH, Redox, rH:
BNC-socket
Temperature:
4-pin screened Mini-DIN-plug for Pt100 4-wire (2-wire also possible)
Input resistance:
10
12
Ohm (pH, Redox)
Nominal temperature:
25°C
Working temperature:
0 to +50°C
Relative humidity:
0 to +95%RH (non-condensing)
Storage temperature:
-20 to +70°C
Interface:
Serial interface (3.5mm jack), serial interface can be connected to RS232 or USB inter-
face of a PC via electrically isolated interface adapter GRS3100, GRS3105 or
USB3100 (accessories).
Memory:
Min-, Max-value- and Hold memory
Power supply:
9V-battery, type IEC 6F22 (included) or additional d.c.connector (internal pin Ø 1.9 mm)
for external 10.5-12V direct voltage supply.
(suitable power supply: GNG10/3000)
Power consumption:
approx. 3 mA
Low battery warning:
and
’ bAt ’
Automatic-off-function:
Device will be automatically switched off if no key is pressed/no interface communica-
tion takes place for the time of the power-off delay. The power-off delay can be set to
values between 1and 120 min.; it can be completely deactivated.
Housing dimensions:
142 x 71 x 26 mm (L x W x D), impact-resistant ABS plastic housing, membrane key-
board, transparent panel. Front side IP65, integrated pop-up clip for table top or sus-
pended use.
Weight:
approx. 165 g
Temp. compensation:
Automatic temperature compensation (ATC) in the operating modes "pH", "mV
H
" or "rH"
if temperature probe is used. Compensation temp. range: 0 ... 105°C
If no temp. probe is used, temperature can be entered manually.
pH-calibration:
- 2-point or 3-pt. calibration with standard buffers, DIN-buffers, manually entered puffers
- autom. buffer detection, temperature dependence of standard or DIN buffers will be
automatically compensated
- permissible electrode data:
asymmetry: ±55 mV
slope: -62...-45 mV/pH
- sensor evaluation according to calibration result (from 10 to 100%).
Rec. redox electrodes:
Reference system: silver/silver chloride, electrolyte: KCl, 3 mol/l
(use only this type if changing values to hydrogen system "mVH" and for rH measurements!)
EMV:
The device corresponds to the essential protection ratings established in the Regula-
tions of the Council for the Approximation of Legislation for the member countries re-
garding electromagnetic compatibility (2004/108/EG).
Additional fault: <1%