H33.0.02.6C-09 

Operating Manual GMH 3530 

page 14 of 14

 

 

16  Specification: 

Display ranges: 

Temperature:  

-100.0 ... +250.0°C or -148.0 ... +482.0°F 

pH: 

0.00 ... 14.00 pH 

Redox (ORP): 

-1999 ... +2000 mV; 
referring to hydrogen system: -1792 ... +2207 mV

(bei 25°C) acc. to DIN 38404 

rH: 

rH 0.0 ... 70.0 

Resolutions: 

0.1°C or 0.1°F / 0.01 pH / 1 mV / 0.1 rH 

Accuracy:

 (at nominal temperature, device ±1Digit)

 

Temperature: 

±0.2°C (-20..80°C), otherwise ±0.4°C 

pH: 

±0.01 pH (for electrode temperature 10..50°C) 

Redox: 

±0.1% FS (mV or mV

H

rH: 

±0.1 rH 

Sensor connections: 

pH, Redox, rH: 

BNC-socket 

Temperature: 

4-pin screened Mini-DIN-plug for Pt100 4-wire (2-wire also possible) 

Input resistance: 

10

12

 

Ohm  (pH, Redox) 

Nominal temperature: 

25°C 

Working temperature: 

0 to +50°C 

Relative humidity: 

0 to +95%RH (non-condensing) 

Storage temperature: 

-20 to +70°C 

Interface:  

Serial interface (3.5mm jack), serial interface can be connected to RS232 or USB inter-
face of a PC via electrically isolated interface adapter GRS3100, GRS3105 or 
USB3100 (accessories). 

Memory: 

Min-, Max-value- and Hold memory 

Power supply: 

9V-battery, type IEC 6F22 (included) or additional d.c.connector (internal pin Ø 1.9 mm) 
for external 10.5-12V direct voltage supply. 

 

(suitable power supply: GNG10/3000) 

Power consumption: 

approx. 3 mA 

Low battery warning:  

 and 

’ bAt ’ 

Automatic-off-function:  

Device will be automatically switched off if no key is pressed/no interface communica-
tion takes place for the time of the power-off delay. The power-off delay can be set to 
values between 1and 120 min.; it can be completely deactivated. 

Housing dimensions:  

142 x 71 x 26 mm (L x W x D), impact-resistant ABS plastic housing, membrane key-
board, transparent panel. Front side IP65, integrated pop-up clip for table top or sus-
pended use. 

Weight:  

approx. 165 g 

Temp. compensation: 

Automatic temperature compensation (ATC) in the operating modes "pH", "mV

H

" or "rH" 

if temperature probe is used. Compensation temp. range: 0 ... 105°C  
If no temp. probe is used, temperature can be entered manually. 

pH-calibration:  

- 2-point or 3-pt. calibration with standard buffers, DIN-buffers, manually entered puffers 
- autom. buffer detection, temperature dependence of standard or DIN buffers will be 
  automatically compensated 
- permissible electrode data:  

asymmetry: ±55 mV 

  

slope: -62...-45 mV/pH 

- sensor evaluation according to calibration result (from 10 to 100%). 

Rec. redox electrodes: 

Reference system: silver/silver chloride, electrolyte: KCl, 3 mol/l 

(use only this type if changing values to hydrogen system "mVH" and for rH measurements!) 

EMV:

  

The device corresponds to the essential protection ratings established in the Regula-
tions of the Council for the Approximation of Legislation for the member countries re-
garding electromagnetic compatibility (2004/108/EG). 
Additional fault: <1% 

17  Disposal instruction: 

Batteries must not be disposed in the regular domestic waste but at the designated collecting points.  
The device must not be disposed in the unsorted municipal waste! Send the device directly to us (sufficiently 
stamped), if it should be disposed. We will dispose the device appropriate and environmentally sound. 

 

_

+

Содержание GMH 3530

Страница 1: ...09 GREISINGER electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for pH Redox temperature as of version V2 6 Operating Manu...

Страница 2: ...AND SYSTEM MESSAGES 10 13 THE PH ELECTRODE 11 13 1 VARIOUS APPLICATIONS REQUIRE SPECIAL ELECTRODES 11 14 THE REDOX ELECTRODE 12 15 THE SERIAL INTERFACE 13 15 1 THE FOLLOWING INTERFACE FUNCTIONS WILL B...

Страница 3: ...s connected or the connection cable is damaged the display will nevertheless show mV pH or rH values Please note that these values can never be correct measuring results 4 Safety Requirements This dev...

Страница 4: ...ed press for 2 sec the min or max value will be deleted Configuration to enter values or change settings CAL for pH mode only press shortly display state of electrode condition and calibration data pr...

Страница 5: ...for pH measurements only 2 Pt the pH electrode will be calibrated at 2 points one calibration point in neutral range and one calibration point in acid or basic range 3 Pt the pH electrode will be cal...

Страница 6: ...measuring mode 8 Manual Setting of pH Value for measuring function rH Running the GMH 3530 in the rH measuring mode requires both temperature and pH value Press shortly again to switch over to the pH...

Страница 7: ...mum calibration of the device 9 3 How to carry out calibration Please note the calibration can only carried out at a temperature range of 0 60 C If you have not yet done so set device to measuring mod...

Страница 8: ...of temperature for solution 2 Use or keys to enter the buffer solution temperature Use to take over value and to display electrode state 5 Rinse electrode in distilled or non ionised water dry electro...

Страница 9: ...ermine the rH value of your solution proceed as follows Please note Make sure that pH and redox electrodes are in a perfect condition during measuring and that they are cleaned and dried carefully bef...

Страница 10: ...eck are there any values below the measur ing range specified meas device not suitable Electrode sensor cable defective replace electrode probe System fault switch on again if fault continues to exist...

Страница 11: ...embrane or the diaphragm will influence the measurements and have to be removed at regular intervals This can be done by means of automatic cleaning equipment 13 1 Various applications require special...

Страница 12: ...mometer Check reference electrolyte level at regular intervals if necessary top up electrolyte level with 3 mol lKCI solution through filling hole closed by means of a silicon ring using a syringe or...

Страница 13: ...ting mode pH mV rH Ther Channel 1 2 1 2 1 2 1 x x x x x x x 0 Read nominal value x x x 1 Set nominal value x x x x x x x 3 Read system status x x x x x x x 6 Read min value x x x x x x x 7 Read max va...

Страница 14: ...elay can be set to values between 1and 120 min it can be completely deactivated Housing dimensions 142 x 71 x 26 mm L x W x D impact resistant ABS plastic housing membrane key board transparent panel...

Отзывы: