H33.0.02.6C-09 

Operating Manual GMH 3530 

page 11 of 14

 

 

 

Slope is too high 
- Electrode defective 
- Buffer solution defective 

 
replace electrode 
use new buffer solution 

 

Incorrect calibration temperature 

calibration can only be done at 0..60°C 

13  The pH-electrode 

pH-electrodes are wear parts which need to be replaced, depending on the chemical or mechanical stress they are 
subjected to, if the values required can no longer be kept even after thorough cleaning and recovery. Please take 
into account that there are several materials which attack glass when they are in water solutions; other chemicals 
may react with the KCl-solution in the electrode thus causing blockings in the diaphragm.  

Examples:  

-  with solutions containing protein, like they are used on the medical and biological sector, KCI may result in the 

denaturation of the protein. 

-  coagulated varnish 
-  solutions with a relatively high concentration of silver ions 

Problems may also occur when taking measurements in low-ion media containing solvent. Some of the problems 
occurring when taking measurements in such media can be counteracted by using a double-chamber electrode 

(Type GE 103) 

with suitable bridging electrolyte (type depending on application). 

 

Any material depositing on the measuring membrane or the diaphragm will influence the measurements and have to 
be removed at regular intervals. This can be done by means of automatic cleaning equipment. 

13.1  Various applications require special electrodes 

1. 

Measurements in low-ion media

 (rain water, aquarium water, VE-waters) 

Type GE 104 (special facetted electrode as of 50µS/cm) or GE 106 (as of 100µS/cm)

2. 

Sea water aquariums 

Standard pH combined electrodes with 3mol KCl 

(type GE 100)

.

 

3. 

Photographic laboratories 

Use double-chamber electrodes, with bridging electrolyte (1 molar potassium-nitrate solution); potassium-nitrate 
solution has to be exchanged if necessary, make sure to fill water cap for storage of electrode with potassium-
nitrate solution 

(type GE 103)

4. 

Pools 

Standard pH-electrode with 3mol KCl 

(type GE 100 )

.

 

5. 

Soil checks 

Glass electrode with several diaphragms 

(type GE 101)

. Use insertion mandrel! 

6. 

Cheese, fruit, meat 

Insertion electrode 

(type GE 101). 

When taking measurements in cheese, milk and other high-protein products 

use special cleaning agent to clean electrode. 

(pepsin solution - GRL 100)

.  

Standard cleaning:

 apply 0,1 molar HCl-solution for at least 5 minutes or protein cleaning agent. 

The average service life of an electrode is 8 to 10 months but may be increased to 2 years if electrode is well main-
tained and treated carefully. We regret not being able to give a more detailed information as this is highly dependent 
on the individual case of application. 

 

m b a r

m b a r

m b a r

m b a r

Содержание GMH 3530

Страница 1: ...09 GREISINGER electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for pH Redox temperature as of version V2 6 Operating Manu...

Страница 2: ...AND SYSTEM MESSAGES 10 13 THE PH ELECTRODE 11 13 1 VARIOUS APPLICATIONS REQUIRE SPECIAL ELECTRODES 11 14 THE REDOX ELECTRODE 12 15 THE SERIAL INTERFACE 13 15 1 THE FOLLOWING INTERFACE FUNCTIONS WILL B...

Страница 3: ...s connected or the connection cable is damaged the display will nevertheless show mV pH or rH values Please note that these values can never be correct measuring results 4 Safety Requirements This dev...

Страница 4: ...ed press for 2 sec the min or max value will be deleted Configuration to enter values or change settings CAL for pH mode only press shortly display state of electrode condition and calibration data pr...

Страница 5: ...for pH measurements only 2 Pt the pH electrode will be calibrated at 2 points one calibration point in neutral range and one calibration point in acid or basic range 3 Pt the pH electrode will be cal...

Страница 6: ...measuring mode 8 Manual Setting of pH Value for measuring function rH Running the GMH 3530 in the rH measuring mode requires both temperature and pH value Press shortly again to switch over to the pH...

Страница 7: ...mum calibration of the device 9 3 How to carry out calibration Please note the calibration can only carried out at a temperature range of 0 60 C If you have not yet done so set device to measuring mod...

Страница 8: ...of temperature for solution 2 Use or keys to enter the buffer solution temperature Use to take over value and to display electrode state 5 Rinse electrode in distilled or non ionised water dry electro...

Страница 9: ...ermine the rH value of your solution proceed as follows Please note Make sure that pH and redox electrodes are in a perfect condition during measuring and that they are cleaned and dried carefully bef...

Страница 10: ...eck are there any values below the measur ing range specified meas device not suitable Electrode sensor cable defective replace electrode probe System fault switch on again if fault continues to exist...

Страница 11: ...embrane or the diaphragm will influence the measurements and have to be removed at regular intervals This can be done by means of automatic cleaning equipment 13 1 Various applications require special...

Страница 12: ...mometer Check reference electrolyte level at regular intervals if necessary top up electrolyte level with 3 mol lKCI solution through filling hole closed by means of a silicon ring using a syringe or...

Страница 13: ...ting mode pH mV rH Ther Channel 1 2 1 2 1 2 1 x x x x x x x 0 Read nominal value x x x 1 Set nominal value x x x x x x x 3 Read system status x x x x x x x 6 Read min value x x x x x x x 7 Read max va...

Страница 14: ...elay can be set to values between 1and 120 min it can be completely deactivated Housing dimensions 142 x 71 x 26 mm L x W x D impact resistant ABS plastic housing membrane key board transparent panel...

Отзывы: