H72.0.01.6C-02
Operating Manual GMH 3431
page 6 of 14
_____________________________________________________
_____________________________________________________________________________
4.4 Pop-up clip
Handling:
‐
Pull at label “open” in order to swing open the pop-up clip.
‐
Pull at label “open” again to swing open the pop-up clip further.
Pop-up clip closed
Pop-up clip at position 90°
Pop-up clip at position 180°
Function:
‐
The device with a closed pop-up clip can be plainly laid onto a table or attached to a belt, etc.
‐
The device with pop-up clip at position 90° can be set up on a table, etc.
‐
The device with pop-up clip at position 180° can be suspended from a screw or the magnetic holder
GMH 1300.
Device
attached
to
a
belt
Device
set
up
on
a
table
Device
suspended
from
magnetic
holder
GMH
1300
5 Start
Operation
Turn device on via
key.
After segment test
the device displays some information on its configuration:
if cell correction scale was changed (cell correction scale unequal 1.000)
(see chapter 10 Automatic adjustment/calibration of cell correction)
if zero point or slope correction is active
(see chapter 9 Adjustment of temperature input)
After that the device is ready for measuring.