GREISINGER GGO Series Скачать руководство пользователя страница 6

Z11.0.0X.6C-21 

Operating manual 

GGO -, GGA -, GOO -, GOG -

 

  page 6 of 8 

3. Application of the different sensor types  

3.1. GGO housing (closed sensor) 

For measurements at atmosphere and in systems without 
over or under pressure the GGO... is sufficient. Additionally 
the GGO can be screwed tightly into systems with small 
over or under pressure. Attention! Mind the maximum pres-
sure and the maximum pressure difference at the mem-
brane.

 

GMH 3690: Always calibrate and measure at the same pressure!  
GMH 3691: If pressure during calibration and measuring differs, please enter its value manually. 
GMH 3692/3695: if instrument and sensor pressure are identical, it will be compensated  
automatically. 

3.2. GOO housing (open sensor) 

The sensor is equipped with drillings at the end and be-
cause of its special construction the measuring gas 
streams optimally around the sensor. No pressure can 
appear while gas blows to the sensor, which otherwise 
would result in erroneous measures. The temperature 
compensation speed of the sensor also is optimised by 

this design. 

Especially the measuring of gases from compressed gas bottles

, where the expansion of the 

gas leaving the bottle lowers the temperature, is optimised with regard to the temperature compen-
sation and pressure errors. The gas flow should be chosen in a suitable range, where no overpres-
sure can happen, esp. if the sensor is connected directly to the source e.g. by means of a tube.

 

3.3. GGA housing (closed sensor with pressure connection) 

If You use an instrument with pressure connection 
(e.g. GMH 3695), this type is optimal. Especially at 

systems with over or under pressure

 or at dy-

namic pressure due to gas flow. It can be screwed 
tightly into systems with small over or under pres-
sure. Attention! Mind the maximum pressure. The 
instrument pressure port is connected directly to the 

sensor port. Then the actual pressure at the membrane will be measured and compensated auto-
matically. 

3.4. GOG housing (ResOx Measuring) 

The GOG / ResOx housings are specialized gas 
sampling devices, which are suitable to sample 
and analyse small gasprobes by means of a gas 
sampling pump.  
You get more informations in the referring GOG 
und ResOx set manuals 

 

4. Measuring Oxygen with the GMH 369x and GMH 569x 

Calibration and measuring are depending of the absolute pressure at the sensor.

 

Therefore check the absolute pressure before calibration and measuring. When using a GMH 3691 
(with manual pressure compensation) the pressure can be entered, when using a GMH 3690 the 
pressure during calibration should be the same as during the measuring (otherwise e.g. 1% devia-
tion results in 1% measuring error).

 

The GMH 3692 /3695 /5690 / 5695 have integrated automatic pressure compensation. There you 
only have to have in mind, that instrument and sensor membrane have the same pressure. 

Sensor temperature and gas temperature should be the same.

 

Temperature differences may cause additional measuring errors! In worst case conditions it may 
take up to several hours until both temperatures are adjusted. A suitable flow of the gas around the 
sensor element increases the adjustment significantly. 

Содержание GGO Series

Страница 1: ...NY 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Operating Manual Oxygen partial pressure sensor in specialized housing GGO GGA GOO GOG ResOx Please read these instructions carefully before us...

Страница 2: ...ly read and understood this manual before starting any work The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one ignoring advices of this manu...

Страница 3: ...age to the device or environment at non observance Note This symbol point out processes which can indirectly influence operation pos sibly cause incorrect measurement or provoke unforeseen reactions a...

Страница 4: ...has adjusted to the ambient temperature before trying a new start up 7 The sensor contains KOH GOEL 381 or acid GOEL 370 KOH and acid can cause severe chemical burns If leaking avoid contact If there...

Страница 5: ...are available which happens normally at the end of life time The nominal life time can be shortened significantly by usage Influencing factors are Storage Operation temperature Humidity of measured ga...

Страница 6: ...th pressure connection If You use an instrument with pressure connection e g GMH 3695 this type is optimal Especially at systems with over or under pressure or at dy namic pressure due to gas flow It...

Страница 7: ...crew in sensor 3 in part 1 Do not forget O Ring 2 Unscrew the sensor carefully e g by means of suitable nippers The audio plug of part 4 has to be connected to the socket in the sensor If this makes p...

Страница 8: ...rolyte basic acidic Connection approx 1 m cable Temperature sensor integrated Storage temperature 15 to 60 C Operating ambient 0 45 C 0 50 C Ambient pressure 0 6 1 75 bar abs 0 5 to 2 0 bar abs Over u...

Отзывы: