GREISINGER GGO Series Скачать руководство пользователя страница 2

Z11.0.0X.6C-21 

Operating manual 

GGO -, GGA -, GOO -, GOG -

 

  page 2 of 8 

Contents 

1. SAFETY ....................................................................................................................................................... 2

 

1.1. General note ............................................................................................................................................ 2

 

1.2. Intended use ............................................................................................................................................ 2

 

1.3. Skilled personnel ..................................................................................................................................... 3

 

1.4. Safety signs and symbols ........................................................................................................................ 3

 

1.5. Reasonably foreseeable misuse ............................................................................................................... 3

 

1.6. Safety guidelines ..................................................................................................................................... 4

 

2. PRODUCT DESCRIPTION ....................................................................................................................... 4

 

2.1. Scope of supply ....................................................................................................................................... 4

 

2.2. General information about application areas .......................................................................................... 4

 

2.3. General information about the oxygen sensors ....................................................................................... 5

 

3. APPLICATION OF THE DIFFERENT SENSOR TYPES .................................................................... 6

 

3.1. GGO  (closed sensor) .............................................................................................................................. 6

 

3.2. GOO (open sensor) ................................................................................................................................. 6

 

3.3. GGA (closed sensor with pressure connection) ...................................................................................... 6

 

4. MEASURING OXYGEN WITH THE GMH 369X AND GMH 569X ................................................... 6

 

5. HOW TO OPERATE .................................................................................................................................. 7

 

6. EXCHANGING THE SENSOR ELEMENT ............................................................................................ 7

 

7. DECOMMISSIONING, RESHIPMENT AND DISPOSAL.................................................................... 7

 

7.1. Decommissioning ................................................................................................................................... 7

 

7.2. Reshipment and disposal ......................................................................................................................... 7

 

8. SPECIFICATION ....................................................................................................................................... 8

 

1. Safety  

1.1. General note 

Read this document carefully and get used to the operation of the device before you use it.  
Keep this document within easy reach near the device for consulting in case of doubt. 

Mounting, start-up, operating, maintenance and removing from operation must be done by quali-
fied, specially trained staff that have carefully read and understood this manual before starting any 
work.  

The manufacturer will assume no liability or warranty in case of usage for other purpose than the 
intended one, ignoring advices of this manual, operating by unqualified staff as well as unauthor-
ized modifications to the device. 

1.2. Intended use 

The sensor is suitable for the measuring of gaseous oxygen, in conjunction with appropriate de-
vices. 

Dependent on the application, a suitable sensor-element must be chosen.

  

See also the specification at the end of the document. 

Personnel which starts up, operates and maintains the device has to have sufficient knowledge of 
the measuring procedure and the meaning of the resulting measured values, this manual delivers a 
valuable help for this. The instructions of the manual have to be understood, regarded and fol-
lowed.  

To be sure that there´s no risk arising due to misinterpretation of measured values, the operator 
must have further knowledge in case of doubt - the user is liable for any harm/damage resulting 
from misinterpretation due to insufficient knowledge. 

The manufacturer will assume no liability or warranty in case of usage for other purpose than the 
intended one, ignoring this manual, operating by unqualified staff as well as unauthorized modifica-
tions to the device. 
 

 

Содержание GGO Series

Страница 1: ...NY 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Operating Manual Oxygen partial pressure sensor in specialized housing GGO GGA GOO GOG ResOx Please read these instructions carefully before us...

Страница 2: ...ly read and understood this manual before starting any work The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one ignoring advices of this manu...

Страница 3: ...age to the device or environment at non observance Note This symbol point out processes which can indirectly influence operation pos sibly cause incorrect measurement or provoke unforeseen reactions a...

Страница 4: ...has adjusted to the ambient temperature before trying a new start up 7 The sensor contains KOH GOEL 381 or acid GOEL 370 KOH and acid can cause severe chemical burns If leaking avoid contact If there...

Страница 5: ...are available which happens normally at the end of life time The nominal life time can be shortened significantly by usage Influencing factors are Storage Operation temperature Humidity of measured ga...

Страница 6: ...th pressure connection If You use an instrument with pressure connection e g GMH 3695 this type is optimal Especially at systems with over or under pressure or at dy namic pressure due to gas flow It...

Страница 7: ...crew in sensor 3 in part 1 Do not forget O Ring 2 Unscrew the sensor carefully e g by means of suitable nippers The audio plug of part 4 has to be connected to the socket in the sensor If this makes p...

Страница 8: ...rolyte basic acidic Connection approx 1 m cable Temperature sensor integrated Storage temperature 15 to 60 C Operating ambient 0 45 C 0 50 C Ambient pressure 0 6 1 75 bar abs 0 5 to 2 0 bar abs Over u...