background image

G 1720

Safety | 2

B-H86.0.0C.DB2-1.0

7 / 25

 CAUTION

Erratic behaviour!

On suspicion that the product can no longer be operated without danger, it must be
decommissioned and prevented from recommissioning with appropriate labelling. The
safety of the user can be impaired by the device if, for example, if it shows visible
damage, it no longer works as specified or if it was stored for an extended period of
time under unsuitable conditions.

– Visual inspection!

– In case of doubt, send the product to the manufacturer for repair or maintenance!

 CAUTION

Stab injury!

Products with insertion probes entail the risk of stab injuries due to the pointed probe
design.

– Handle insertion probes with care!

– Fit a protective cap on the measuring probe!

 NOTE

If the product is stored at a temperature above 50 °C, or is not used for an extended
period of time, the batteries must be removed. Leaks from the batteries are avoided as
a result.

 NOTE

This product does not belong in children's hands!

 NOTE

The sensor handle, connecting cable and product housing are not designed for con-
tinuous contact with foods.

Designed for continuous contact with foods in accordance with EC Regulation 1935 /
2004:

– The temperature sensor from the measuring tip to approx. 1 cm before the end of

the stainless steel tube.

2.4 Intended use

The product is a water-protected thermometer. It is designed for precise and instant-
aneous temperature measurements in the following media:

– Food

– Liquids

– Gases

– Soft plastic materials

– Bulk material

See 

Technical data [

}

 p. 22]

.

Содержание G 1720

Страница 1: ...Operating manual G 1720 Alarm thermometer Permanently connected sensor Waterproof Precise and fast EN...

Страница 2: ...ments 10 5 Operation 11 5 1 Commissioning 11 5 1 1 Explanation 11 5 2 Configuration 11 5 2 1 Explanation 11 5 2 2 Opening the configuration menu 11 5 2 3 Configuring parameters of the configuration me...

Страница 3: ...Table of contents B H86 0 0C DB2 1 0 3 25 9 Disposal 21 10 Technical data 22 11 Spare parts and accessories 23 12 Service 24 12 1 Manufacturer 24 12 2 Repairs 24 12 3 Sales subsidiaries 25...

Страница 4: ...Legal notices The liability and warranty of the manufacturer for damages and consequential dam ages are voided with misuse disregarding this operating manual disregarding safety notices assignment of...

Страница 5: ...duct Outcome of an action Result consequence or effect of an instruction Emphases In order to simplify legibility and provide a clearer overview various sections informa tion are emphasised 1234 Displ...

Страница 6: ...can only be guaranteed if generally applicable safety precautions and the device specific safety instructions for this document are observed If these notices are disregarded personal injury or death...

Страница 7: ...le insertion probes with care Fit a protective cap on the measuring probe NOTE If the product is stored at a temperature above 50 C or is not used for an extended period of time the batteries must be...

Страница 8: ...which purpose this document makes a valuable contribution The instructions in this docu ment must be understood observed and followed In order to ensure that no risks arise from the interpretation of...

Страница 9: ...l impressive features include the dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which offers overhead display at the push of a button The product can...

Страница 10: ...easurement of the current temperature in min max hold mode with unit 4 3 Operating elements On Off button Press briefly Switch on the product Activate deactivate lighting Long press Switch off the pro...

Страница 11: ...roduct for your purposes NOTE There are various configuration parameters available depending on the product ver sion and configuration They can differ depending on the product version and configur ati...

Страница 12: ...d parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up button and Down button 3 The available configuration options are listed for each parame...

Страница 13: ...ted with the zero point correction and the gradient correction If an adjustment is made you change the pre adjusted factory settings This is signalled with the T OF or T SL when the product is switche...

Страница 14: ...ired parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up button and Down button 3 The available configuration options are listed for each par...

Страница 15: ...o point T OF 0 2 C 0 0 C 0 2 C T SL setpoint gradient correction display gradient correction T OF 1 100 T SL 37 0 C 36 6 C 0 2 1 100 0 54 Outcome of an action The changed value is saved and the Config...

Страница 16: ...ntal temperature influ ence the measurement result NOTE Thermally conductive paste between the measuring sensor and surface can also in crease measurement accuracy in some cases 6 1 3 Cooling evaporat...

Страница 17: ...ate for a certain length of time If the BAT display text appears in the main display the battery voltage is no longer ad equate for operation of the product Now the battery is fully depleted 7 2 2 Cha...

Страница 18: ...are categorised as ISO calibration certificates and DAkkS calibration certificates The purpose of the calibration is to verify the precision of the measuring device by comparing it with a traceable re...

Страница 19: ...0 0C DB2 1 0 19 25 NOTE The product is delivered with a test report This confirms that the measuring device has been adjusted and tested NOTE Only the manufacturer can check the basic settings and mak...

Страница 20: ...m error Product is defective Battery depleted Error in the product Product is defective Replace battery Send in for repair BAT Battery depleted Battery depleted Replace battery ERR 1 Measuring range e...

Страница 21: ...se online at www ghm group de and sent it in with the product NOTE The device must not be disposed of with household waste If the product is disposed of please take it to a municipal collection point...

Страница 22: ...ut kink protection Weight 150 g incl battery and sensor Operating conditions 20 to 50 C 0 to 95 r h temporarily 100 r h Storage temperature 20 to 70 C Current supply 2 AA battery included in the scope...

Страница 23: ...ed below Article Number Name Description 610049 Mignon battery AA Mignon AA spare battery 611373 ST G1000 Device protection bag with 1 round cut out A complete list of all accessories and spare parts...

Страница 24: ...nger de WEEE reg no DE 93889386 12 2 Repairs Defective products are repaired professionally and quickly in our service centre Open hours and contact Monday to Thursday from 8 00 to 16 00 Friday from 8...

Страница 25: ...UP France SAS Parc des Pivolles 9 Rue de Catalogne 69150 D cines Charpieu Lyon FRANCE Phone 33 4 72 37 45 30 a jouanilou ghm group fr India GHM Messtechnik India Pvt Ltd 209 Udyog Bhavan Sonowala Road...

Отзывы: